Performance comparison of 4 portable oxygen concentrators.

Robert L Chatburn1, Thomas J Williams.   

Abstract

BACKGROUND: Several portable oxygen concentrators (POCs) have become available in the United States. Technical trade-offs are made in designing POCs, so their performance is expected to differ.
METHODS: We tested 4 POC models (Invacare XPO(2), Respironics EverGo, AirSep FreeStyle, and Inogen One) for oxygen delivery as a function of respiratory rate. We measured oxygen volume per pulse, pulse duration, trigger sensitivity, oxygen concentration in the gas delivered by the POC (measured with a ceramic-based oxygen sensor), and relative fraction of inspired oxygen (relative F(IO(2))), measured with a setup that included an adult nasal cannula, a model nares, and a lung simulator that has a built-in oxygen sensor. We studied respiratory rates of 15, 20, 25, 30, and 35 breaths/min, at a tidal volume of 500 mL.
RESULTS: The XPO(2) had the highest pulse flow, and FreeStyle had the lowest, which corresponded to the highest and lowest pulse volumes (40.8 mL and 18.2 mL respectively) at POC setting 2, at 15 breaths/min. The range of oxygen purity was 91.9-94.4% (P = .36). The Inogen had the shortest pulse-delivery time, and the XPO(2) had the longest (132 ms vs 281 ms, P < .001). The FreeStyle had the highest trigger sensitivity, and the Inogen the lowest (0.15 cm H(2)O vs 0.21 cm H(2)O, P < .001). At the maximum settings, with all 4 POCs, relative F(IO(2)) decreased as respiratory rate increased (relative F(IO(2)) range 28.6-31.4% at 15 breaths/min, 23.0-25.3% at 30 breaths/min).
CONCLUSIONS: These 4 POC models have markedly different performance, which emphasizes the need to adjust the POC setting to meet the specific patient's needs at rest and with activity.

Entities:  

Mesh:

Year:  2010        PMID: 20406511

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  11 in total

1.  Accuracy of a novel system for oxygen delivery to small children.

Authors:  Matthew Coghill; Namasivayam Ambalavanan; Robert L Chatburn; Patricia L Hibberd; Linda L Wright; Waldemar A Carlo
Journal:  Pediatrics       Date:  2011-07-04       Impact factor: 7.124

2.  Variability in uptake efficiency for pulsed versus constant concentration delivery of inhaled nitric oxide.

Authors:  Andrew R Martin; Chris Jackson; Ira M Katz; Georges Caillibotte
Journal:  Med Gas Res       Date:  2014-01-22

3.  Brain adaptation to hypoxia and hyperoxia in mice.

Authors:  Laura Terraneo; Rita Paroni; Paola Bianciardi; Toniella Giallongo; Stephana Carelli; Alfredo Gorio; Michele Samaja
Journal:  Redox Biol       Date:  2016-11-04       Impact factor: 11.799

4.  An in silico analysis of oxygen uptake of a mild COPD patient during rest and exercise using a portable oxygen concentrator.

Authors:  Ira Katz; Marine Pichelin; Spyridon Montesantos; Min-Yeong Kang; Bernard Sapoval; Kaixian Zhu; Charles-Philippe Thevenin; Robert McCoy; Andrew R Martin; Georges Caillibotte
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2016-09-29

5.  Contemporary portable oxygen concentrators and diverse breathing behaviours -- a bench comparison.

Authors:  Dion C Martin
Journal:  BMC Pulm Med       Date:  2019-11-19       Impact factor: 3.317

6.  Shedding Light on the Main Characteristics and Perspectives of Romanian Medicinal Oxygen Market.

Authors:  Adriana AnaMaria Davidescu; Simona Andreea Apostu; Cristina Stanciu-Mandruleanu
Journal:  Healthcare (Basel)       Date:  2021-02-03

7.  Flexible oxygen concentrators for medical applications.

Authors:  Akhil Arora; M M Faruque Hasan
Journal:  Sci Rep       Date:  2021-07-12       Impact factor: 4.379

8.  Comparison of pulsed versus continuous oxygen delivery using realistic adult nasal airway replicas.

Authors:  John Z Chen; Ira M Katz; Marine Pichelin; Kaixian Zhu; Georges Caillibotte; Michelle L Noga; Warren H Finlay; Andrew R Martin
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2017-08-24

9.  The Impact of Moderate Chronic Hypoxia and Hyperoxia on the Level of Apoptotic and Autophagic Proteins in Myocardial Tissue.

Authors:  Alexandra Gyongyosi; Laura Terraneo; Paola Bianciardi; Arpad Tosaki; Istvan Lekli; Michele Samaja
Journal:  Oxid Med Cell Longev       Date:  2018-08-16       Impact factor: 6.543

Review 10.  Oxygen Supply in Hospitals: Requisites in the Current Pandemic.

Authors:  Ragi Jain; Charu Sharma
Journal:  Anesth Essays Res       Date:  2022-02-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.