Value and limitations of transesophageal echocardiography in assessment of mitral valve prostheses.

B K Khandheria1, J B Seward, J K Oh, W K Freeman, B A Nichols, L J Sinak, F A Miller, A J Tajik.   

Abstract

BACKGROUND: Transthoracic Doppler echocardiography examination has become an integral part of the investigations performed in patients with mitral valve prostheses. The limitations of the transthoracic approach are well documented. Transesophageal echocardiography provides a unique window for achieving a clear view of the mitral prosthesis. METHODS AND
RESULTS: This study shows the usefulness of transesophageal echocardiography in clinical practice for assessment of patients with a mitral valve prosthesis. This technique demonstrated an abnormality in 48% of patients who had normal results on transthoracic examination. The overall sensitivity of transesophageal echocardiography was 96%.
CONCLUSIONS: Transesophageal echocardiography constitutes an essential part of a comprehensive two-dimensional/Doppler echocardiographic examination in patients with suspected malfunction of mitral prostheses.

Entities:  

Mesh:

Year:  1991        PMID: 2040048     DOI: 10.1161/01.cir.83.6.1956

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  15 in total

Review 1.  Transoesophageal echocardiography.

Authors:  J E Sanderson; W W Chan
Journal:  Postgrad Med J       Date:  1997-03       Impact factor: 2.401

Review 2.  Echocardiography in infective endocarditis.

Authors:  A Evangelista; M T Gonzalez-Alujas
Journal:  Heart       Date:  2004-06       Impact factor: 5.994

3.  Retrospective studies of left atrial thrombus: does misclassification impair their clinical utility?

Authors:  D J Ballard
Journal:  J Gen Intern Med       Date:  1991 Mar-Apr       Impact factor: 5.128

Review 4.  Transoesophageal echocardiography.

Authors:  Partho P Sengupta; Bijoy K Khandheria
Journal:  Heart       Date:  2005-04       Impact factor: 5.994

5.  Perivalvular leakage after termination of cardiopulmonary bypass procedure.

Authors:  Y Kadoi; H Kawahara; N Fujita
Journal:  J Anesth       Date:  1997-06       Impact factor: 2.078

6.  Successful thrombolysis of mechanical mitral valve prosthesis in a patient with cardiogenic shock.

Authors:  Christine S Zuern; Tobias Hoevelborn; Anette Wisbar; Martin H Hust; Karl K Haase; Christian Herdeg; Meinrad Gawaz; Andreas E May
Journal:  BMJ Case Rep       Date:  2009-04-14

7.  Prevalence and clinical significance of incidental paraprosthetic valvar regurgitation: a prospective study using transoesophageal echocardiography.

Authors:  A Ionescu; A G Fraser; E G Butchart
Journal:  Heart       Date:  2003-11       Impact factor: 5.994

8.  Cardiac computed tomography angiography results in diagnostic and therapeutic change in prosthetic heart valve endocarditis.

Authors:  Jesse Habets; Wilco Tanis; Lex A van Herwerden; Renee B A van den Brink; Willem P Th M Mali; Bas A J M de Mol; Steven A J Chamuleau; Ricardo P J Budde
Journal:  Int J Cardiovasc Imaging       Date:  2013-11-30       Impact factor: 2.357

9.  Additional value of biplane transoesophageal imaging in assessment of mitral valve prostheses.

Authors:  K Groundstroem; D Rittoo; P Hoffman; P Bloomfield; G R Sutherland
Journal:  Br Heart J       Date:  1993-09

10.  Assessment of mitral prosthetic valve thrombosis by live three-dimensional echocardiography: comparison with transesophageal echocardiography.

Authors:  Antonella Moreo; Elisabetta Lobiati; Benedetta De Chiara; Francesco Mauri