Influence of different contributions of scatter and attenuation on the threshold values in contrast-based algorithms for volume segmentation.

Roberta Matheoud1, Patrizia Della Monica, Chiara Secco, Gianfranco Loi, Marco Krengli, Eugenio Inglese, Marco Brambilla.   

Abstract

The aim of this work is to evaluate the role of different amount of attenuation and scatter on FDG-PET image volume segmentation using a contrast-oriented method based on the target-to-background (TB) ratio and target dimensions. A phantom study was designed employing 3 phantom sets, which provided a clinical range of attenuation and scatter conditions, equipped with 6 spheres of different volumes (0.5-26.5 ml). The phantoms were: (1) the Hoffman 3-dimensional brain phantom, (2) a modified International Electro technical Commission (IEC) phantom with an annular ring of water bags of 3 cm thickness fit over the IEC phantom, and (3) a modified IEC phantom with an annular ring of water bags of 9 cm. The phantoms cavities were filled with a solution of FDG at 5.4 kBq/ml activity concentration, and the spheres with activity concentration ratios of about 16, 8, and 4 times the background activity concentration. Images were acquired with a Biograph 16 HI-REZ PET/CT scanner. Thresholds (TS) were determined as a percentage of the maximum intensity in the cross section area of the spheres. To reduce statistical fluctuations a nominal maximum value is calculated as the mean from all voxel > 95%. To find the TS value that yielded an area A best matching the true value, the cross section were auto-contoured in the attenuation corrected slices varying TS in step of 1%, until the area so determined differed by less than 10 mm² versus its known physical value. Multiple regression methods were used to derive an adaptive thresholding algorithm and to test its dependence on different conditions of attenuation and scatter. The errors of scatter and attenuation correction increased with increasing amount of attenuation and scatter in the phantoms. Despite these increasing inaccuracies, PET threshold segmentation algorithms resulted not influenced by the different condition of attenuation and scatter. The test of the hypothesis of coincident regression lines for the three phantoms used provided no statistical basis for believing that the three lines are not coincident. Calibration curves needed to implement contouring algorithms based on adaptive TS segmentation of PET volumes can be devised in different conditions of attenuation and scatter. This opens the possibility of defining a unified contrast-based method for target delineation in different anatomical districts.
© 2010 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20399128     DOI: 10.1016/j.ejmp.2010.02.003

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

Review 1.  A review on segmentation of positron emission tomography images.

Authors:  Brent Foster; Ulas Bagci; Awais Mansoor; Ziyue Xu; Daniel J Mollura
Journal:  Comput Biol Med       Date:  2014-04-28       Impact factor: 4.589

2.  An Adaptive Thresholding Method for BTV Estimation Incorporating PET Reconstruction Parameters: A Multicenter Study of the Robustness and the Reliability.

Authors:  M Brambilla; R Matheoud; C Basile; C Bracco; I Castiglioni; C Cavedon; M Cremonesi; S Morzenti; F Fioroni; M Giri; F Botta; F Gallivanone; E Grassi; M Pacilio; E De Ponti; M Stasi; S Pasetto; S Valzano; D Zanni
Journal:  Comput Math Methods Med       Date:  2015-05-19       Impact factor: 2.238

3.  Influence of reconstruction settings on the performance of adaptive thresholding algorithms for FDG-PET image segmentation in radiotherapy planning.

Authors:  Roberta Matheoud; Patrizia Della Monica; Gianfranco Loi; Luca Vigna; Marco Krengli; Eugenio Inglese; Marco Brambilla
Journal:  J Appl Clin Med Phys       Date:  2011-01-30       Impact factor: 2.102

Review 4.  The Strategies to Homogenize PET/CT Metrics: The Case of Onco-Haematological Clinical Trials.

Authors:  Stephane Chauvie; Fabrizio Bergesio
Journal:  Biomedicines       Date:  2016-11-15
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.