Literature DB >> 20392748

Application of five frozen human-pooled serum samples assigned by the International Federation of Clinical Chemistry and Laboratory Medicine reference procedure in a traceability investigation of gamma-glutamyltransferase catalytic concentration measurements in China.

Changyu Xia1, Qing Tong, Qingtao Wang, Zhuling Tang, Lili Qi, Shan Chi, Man Zhang, Xuejing Wang, Haixia Li, Guobin Xu.   

Abstract

BACKGROUND: The in vitro directive of the European Union requires traceability to the international recommended reference procedures. The application of the reference procedures is necessary in order to evaluate the accuracy of gamma-glutamyltransferase (GGT) assays of routine measurement systems in China.
METHODS: Five frozen patient-pooled serum samples were assigned values by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference procedure in order to evaluate the traceability of the results of GGT catalytic activity from six homogeneous systems. One of the serum samples was used to calibrate seven non-homogeneous systems.
RESULTS: All of the homogeneous systems, except the Dade system (Dade Bering Inc, IL, USA), achieved traceability within the measurement range. The Roche and Hitachi systems were better than the other systems. After calibration, the variance of the non-homogeneous systems decreased dramatically from between 14.50% and 25.23% to between 1.25% and 3.09% and the bias decreased from between -11.4% and -4.1% to between 0.5% and 3.5%.
CONCLUSION: Manufacturers in China should ensure that their calibration systems correspond to the IFCC reference procedures. Fresh frozen pooled patient serum assigned by reference laboratories can be used to calibrate non-homogeneous systems in order to achieve traceability.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20392748     DOI: 10.1258/acb.2009.009210

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  3 in total

1.  LEGO plot for simultaneous application of multiple quality requirements during trueness verification of quantitative laboratory tests.

Authors:  Hae-il Park; Hyojin Chae; Myungshin Kim; Jehoon Lee; Yonggoo Kim
Journal:  J Clin Lab Anal       Date:  2014-01-06       Impact factor: 2.352

2.  Trueness assessment for serum glucose measurement using commercial systems through the preparation of commutable reference materials.

Authors:  Changyu Xia; Ou Liu; Lanzhen Wang; Guobing Xu
Journal:  Ann Lab Med       Date:  2012-06-20       Impact factor: 3.464

3.  Comparability of four clinical laboratory measurement methods for GGT and commutability of candidate reference materials.

Authors:  Rui Zhang; Qingtao Wang
Journal:  J Clin Lab Anal       Date:  2020-09-11       Impact factor: 3.124

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.