Toe gangrene in an infant subsequent to application of adult-type pulse oximeter probe for 10 min.

Jyotsna Punj1, Ashok Jaryal, Senthil Mahalingam, Chaitra Mukundan, Ravindra Pandey, Vanlal Darlong.   

Abstract

We present a case of ischemic changes after application of an adult pulse oximeter probe in an infant for a short period of only 10 min. To understand the physiology behind this mishap, we studied pressure exerted by the adult pulse oximeter probe on simulated fingers (fluid-filled pouches), which were filled under gravity through the three-way stopcock. A catheter was placed inside the pouch to continuously measure the pressure. The experiment was performed on different sizes of the finger-shaped pouches and by varying the positioning of the pouches in the pulse oximeter. It was observed that pressure exerted by the pulse oximeter was directly proportional to the extent to which it was stretched after placement on the fluid-filled pouches of different sizes. In the fully extended position, the pressure exerted was higher (up to 30 mmHg) than in the fully collapsed position (4 mmHg). Higher pressures were caused by greater stretch of the probe. This could occur due to larger fingers and closer placement of fingers to the hinge of the probe.

Entities:  

Mesh:

Year:  2010        PMID: 20390308     DOI: 10.1007/s00540-010-0930-5

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  9 in total

1.  Pulse oximeter-associated toe injuries in a premature neonate: a case report.

Authors:  C W Lin; H Z Wang; K S Hsieh
Journal:  Zhonghua Yi Xue Za Zhi (Taipei)       Date:  1999-12

2.  Toe gangrene in an infant subsequent to use of pulse oximeter for short duration.

Authors:  Babita Ghai; A Naik; S Rupal; R Madan
Journal:  Anesth Analg       Date:  2005-02       Impact factor: 5.108

3.  Sensory loss of the distal phalanx caused by pulse oximeter probe.

Authors:  M Clark; N G Lavies
Journal:  Anaesthesia       Date:  1997-05       Impact factor: 6.955

4.  Effects of external pressure loading on human skin blood flow measured by 133Xe clearance.

Authors:  G A Holloway; C H Daly; D Kennedy; J Chimoskey
Journal:  J Appl Physiol       Date:  1976-04       Impact factor: 3.531

5.  Ischemic digital skin necrosis: a complication of the reusable Nelcor pulse oximeter probe.

Authors:  K H Berge; W L Lanier; P D Scanlon
Journal:  Anesth Analg       Date:  1988-07       Impact factor: 5.108

6.  Pulse oximeter-induced digital injury: frequency rate and possible causative factors.

Authors:  J Wille; R Braams; W H van Haren; C van der Werken
Journal:  Crit Care Med       Date:  2000-10       Impact factor: 7.598

7.  Skin injury in an infant with pulse oximetry.

Authors:  B Pettersen; U Kongsgaard; H Aune
Journal:  Br J Anaesth       Date:  1992-08       Impact factor: 9.166

8.  Finger injury resulting from pulse oximeter probe during orthognathic surgery.

Authors:  P D Chemello; S R Nelson; L M Wolford
Journal:  Oral Surg Oral Med Oral Pathol       Date:  1990-02

9.  [Skin lesions caused by oxygen saturation monitor probe: pathogenetic considerations concerning 2 neonatal cases].

Authors:  D Trevisanuto; P Ferrarese; F Cantarutti; V Zanardo
Journal:  Pediatr Med Chir       Date:  1995 Jul-Aug