Literature DB >> 20382975

Evaluation of image-enhanced paediatric computed tomography brain examinations.

K Ledenius1, F Stålhammar, L M Wiklund, C Fredriksson, A Forsberg, A Thilander-Klang.   

Abstract

The aim of this study was to evaluate the possibility of reducing the radiation dose to paediatric patients undergoing computed tomography (CT) brain examination by using image-enhancing software. Artificial noise was added to the raw data collected from 20 patients aged between 1 and 10 y to simulate tube current reductions of 20, 40 and 60 mA. All images were created in duplicate; one set of images remained unprocessed whereas the other was processed with image-enhancing software. Three paediatric radiologists assessed the image quality based on their ability to visualise the high- and low-contrast structures and their overall impression of the diagnostic value of the image. For patients aged 6-10 y, it was found that dose reductions from 27 mGy (CTDI(vol)) to 23 mGy (15 %) in the upper brain and from 32 to 28 mGy (13 %) in the lower brain were possible for standard diagnostic CT examinations when using the image-enhancing filter. For patients 1-5 y, the results for standard diagnostics in the upper brain were inconclusive, for the lower brain no dose reductions were found possible.

Entities:  

Mesh:

Year:  2010        PMID: 20382975     DOI: 10.1093/rpd/ncq097

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Reducing radiation exposure from computed tomography of the brain in children--report of a practical approach.

Authors:  Sui-To Wong; Gwendolin Yiu; Yiu-Man Poon; Ming-Keung Yuen; Dawson Fong
Journal:  Childs Nerv Syst       Date:  2012-01-12       Impact factor: 1.475

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.