| Literature DB >> 20382574 |
Kayoko Minakata1, Itaru Yamagishi, Sanae Kanno, Hideki Nozawa, Masako Suzuki, Osamu Suzuki.
Abstract
A rapid and sensitive electrospray ionization (ESI) tandem mass spectrometry (MS-MS) procedure was developed for the determination of iodide (I(-)). A gold (Au) and I(-) complex was formed immediately after the addition of the chelating agent NaAuCl(4) to I(-) solution, and was extracted with methyl isobutyl ketone. One to five microliters of the extract were injected directly into an ESI-MS-MS instrument. I(-) quantification was performed by selecting reaction monitoring of the product ion I(-) at m/z 127 derived from the precursor ion (197)AuI(2)(-) at m/z 451. I(-) concentration was measured in the quantification range from 10(-7) to 10(-5) M using 50 microL of solution within 10 min. Iodate was reduced to I(-) with ascorbic acid and determined. I(-) concentration in reference urine 2670a was measured after treatments. 2010 Elsevier B.V. All rights reserved.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20382574 DOI: 10.1016/j.jchromb.2010.03.022
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205