Multi-slice computed tomography: A tool for non-invasive temperature measurement?

Philipp Bruners1, Elena Levit, Tobias Penzkofer, Peter Isfort, Christina Ocklenburg, Bernhardt Schmidt, Thomas Schmitz-Rode, Rolf W Günther, Andreas H Mahnken.   

Abstract

PURPOSE: To investigate the potential of multi-slice computed tomography (MSCT) as a tool for non-invasive temperature measurement.
MATERIALS AND METHODS: Samples of water, 0.9% saline, sunflower oil and dilutions of (1:32, 1:64, 1:128) contrast agent (Iopromid 370, BayerSchering Pharma, Berlin) were heated in a plexiglass phantom. In a first set-up, samples of 0.9% saline solution were scanned at defined temperatures (25 degrees -75 degrees C; 5 degrees C intervals) using a clinical CT scanner. Scan parameters (tube current-time product, tube voltage, collimation, slice thickness) were systematically varied. In a second set-up samples of the different fluids (water, sunflower oil, contrast agent dilutions) were scanned using the following scan protocol: 250 mAs, 140 kV, 1.2 mm collimation, 9.6 mm slice thickness. CT numbers were measured in reconstructed axial images at the different temperatures. A regression analysis was performed to investigate the relationship between temperature and CT number.
RESULTS: Standard deviation of measured CT numbers decreased with increasing tube current-time product, increasing tube voltage, thicker collimation and higher slice thickness. Regression analysis showed an inverse relationship between temperature and CT number for all fluids with regression coefficients of -0.471 (0.9% saline), 0.447 (water), -0.679 (sunflower oil), -0.420 (contrast agent 1:32), -0.414 (contrast agent 1:64) and -0.441 (contrast agent 1:128), respectively.
CONCLUSION: Multi-slice computed tomography can depict thermal density expansion of different fluids. Based on these results the implementation of a temperature discrimination of several degrees C at a high spatial resolution is achievable.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20377364     DOI: 10.3109/02656731003605654

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  6 in total

Review 1.  Imaging-based internal body temperature measurements: The journal Temperature toolbox.

Authors:  Juho Raiko; Kalle Koskensalo; Teija Sainio
Journal:  Temperature (Austin)       Date:  2020-05-29

2.  Assessment of thermal sensitivity of CT during heating of liver: an ex vivo study.

Authors:  G D Pandeya; M J W Greuter; B Schmidt; T Flohr; M Oudkerk
Journal:  Br J Radiol       Date:  2012-09       Impact factor: 3.039

3.  Feasibility of computed tomography based thermometry during interstitial laser heating in bovine liver.

Authors:  G D Pandeya; J H G M Klaessens; M J W Greuter; B Schmidt; T Flohr; R van Hillegersberg; M Oudkerk
Journal:  Eur Radiol       Date:  2011-03-24       Impact factor: 5.315

Review 4.  Perspectives of breast cancer thermotherapies.

Authors:  Edouard Alphandéry
Journal:  J Cancer       Date:  2014-05-29       Impact factor: 4.207

5.  How large is the periablational zone after radiofrequency and microwave ablation? Computer-based comparative study of two currently used clinical devices.

Authors:  Macarena Trujillo; Punit Prakash; Pegah Faridi; Aleksandar Radosevic; Sergio Curto; Fernando Burdio; Enrique Berjano
Journal:  Int J Hyperthermia       Date:  2020       Impact factor: 3.914

6.  CT-based thermometry with virtual monoenergetic images by dual-energy of fat, muscle and bone using FBP, iterative and deep learning-based reconstruction.

Authors:  Andreas Heinrich; Sebastian Schenkl; David Buckreus; Felix V Güttler; Ulf K-M Teichgräber
Journal:  Eur Radiol       Date:  2021-07-29       Impact factor: 5.315

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.