Literature DB >> 20374683

Vibroplasty involving direct coupling of the floating mass transducer to the oval window niche.

T Zehlicke1, R Dahl, T Just, H-W Pau.   

Abstract

OBJECTIVE: To demonstrate the technique and clinical application of vibroplasty in which the floating mass transducer component of the Vibrant Soundbridge implant is coupled directly to the oval window niche, in patients with a mobile stapes footplate but a malformed or destroyed stapes suprastructure.
METHODS: The underlying concept was to create a soft tissue casing for the floating mass transducer, while also firmly connecting the transducer to a small, solid cartilage 'plunger' attached to the stapes footplate. This was realised by removing almost all the cartilage from a larger piece of cartilage-perichondrium, leaving only a tiny cartilage island about half a millimetre in diameter, attached to a much wider 'blanket' of perichondrium.
RESULTS: By coupling the floating mass transducer directly to the oval window niche, patients' speech understanding was improved. Post-operative aided thresholds of 30-40 dB HL were achieved by all patients.
CONCLUSIONS: In patients with mixed hearing loss combined with a destroyed stapes suprastructure but a mobile stapes footplate, we describe the coupling of the floating mass transducer component of a Vibrant Soundbridge to the stapes footplate, as an alternative to coupling to the round window.

Entities:  

Mesh:

Year:  2010        PMID: 20374683     DOI: 10.1017/S0022215110000526

Source DB:  PubMed          Journal:  J Laryngol Otol        ISSN: 0022-2151            Impact factor:   1.469


  7 in total

1.  Oval and round window vibroplasty: a comparison of hearing results, risks and failures.

Authors:  Andrea Canale; Federico Dagna; Claudia Cassandro; Pamela Giordano; Federico Caranzano; Michelangelo Lacilla; Roberto Albera
Journal:  Eur Arch Otorhinolaryngol       Date:  2013-10-11       Impact factor: 2.503

2.  Vibrant Soundbridge middle ear implantations: experience at National University Hospital Singapore.

Authors:  Lynne Hsueh Yee Lim; Jocelynne Del Prado; Ling Xiang; Abdul Rahman Bin Yusof; Jenny Hooi Yin Loo
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-03-28       Impact factor: 2.503

3.  [The application of implantable hearing aids using the Vibrant Soundbridge as an example].

Authors:  T Strenger; T Stark
Journal:  HNO       Date:  2012-02       Impact factor: 1.284

4.  The Hannover Coupler V2: Audiological outcomes of a round window coupler for the Vibrant Soundbridge.

Authors:  Nicole Knölke; Dawid Murawski; Nina Wardenga; Susan Busch; Hannes Maier; Thomas Lenarz
Journal:  Laryngoscope Investig Otolaryngol       Date:  2022-05-28

5.  Experience with the Vibrant Soundbridge RW-Coupler for round window Vibroplasty with tympanosclerosis.

Authors:  Satoshi Iwasaki; Hiroaki Suzuki; Hideaki Moteki; Maiko Miyagawa; Yutaka Takumi; Shin-Ichi Usami
Journal:  Acta Otolaryngol       Date:  2012-03-04       Impact factor: 1.494

6.  Modified-Power-Piston: Short-Incudial-Process-Vibroplasty and Simultaneous Stapedotomy in Otosclerosis.

Authors:  Daniel Dejaco; David Riedl; Timo Gottfried; Thomas Rasse; Natalie Fischer; David Prejban; Viktor Koci; Herbert Riechelmann; Joachim Schmutzhard; Thomas Keintzel
Journal:  Otol Neurotol       Date:  2019-03       Impact factor: 2.311

7.  Simultaneous Vibrant Soundbridge Implantation and 2nd Stage Auricular Reconstruction for Microtia with Aural Atresia.

Authors:  Lynne Hsueh Yee Lim; Ling Xiang; Jocelynne Del Prado; Ling Ee Chin; Millo Achille Beltrame
Journal:  Audiol Res       Date:  2011-09-23
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.