Multiple-integrating sphere spectrophotometer for measuring absolute spectral reflectance and transmittance.

G A Zerlaut, T E Anderson.   

Abstract

A spectroreflectometer/transmissometer is described that permits determination of absolute optical characteristics in the 300-2600-nm wavelength region (which is essentially the complete solar spectrum). The uniqueness of the instrument derives from use of three rapidly interchangeable 20-cm (8-in.) integrating spheres to measure (1) absolute hemispherical spectral reflectance as a function of angles of incidence from -40 to +40 degrees employing an Edwards-type integrating sphere with a center-mounted sample [using small 2.5-cm (1-in.) diam specimens], (2) absolute hemispherical and absolute diffuse spectral reflectance at an angle of incidence of 20 degrees employing a sphere with a wall-mounted sample (for large specimens) and a screened detector, and (3) absolute hemispherical and absolute directional (near-normal exitance) transmittance employing a complete integrating sphere with the only ports being for the sample and reference beams. Data are presented that demonstrate the ability to measure the spectral reflectance of nonmirror surfaces to an absolute accuracy of 0.995 (an uncertainty of +/-0.005 reflectance units) in both reflectance spheres and of highly specular mirrors to an absolute accuracy of 0.993 (an uncertainty of +/-0.007 reflectance units). Spectral transmittance can be measured to an absolute accuracy of better than 0.995 (an uncertainty of +/-0.005 transmittance units).

Year:  1981        PMID: 20372262     DOI: 10.1364/AO.20.003797

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  3 in total

1.  On the dysfunctional hemoglobins and cyanosis connection: practical implications for the clinical detection and differentiation of methemoglobinemia and sulfhemoglobinemia.

Authors:  Stephen W Askew; Gladimir V G Baranoski
Journal:  Biomed Opt Express       Date:  2018-06-25       Impact factor: 3.732

2.  Lighting Deviation Correction for Integrating-Sphere Multispectral Imaging Systems.

Authors:  Zhe Zou; Hui-Liang Shen; Shijian Li; Yunfang Zhu; John H Xin
Journal:  Sensors (Basel)       Date:  2019-08-10       Impact factor: 3.576

3.  Reflectometers for Absolute and Relative Reflectance Measurements in the Mid-IR Region at Vacuum.

Authors:  Jinhwa Gene; Min Yong Jeon; Sun Do Lim
Journal:  Sensors (Basel)       Date:  2021-02-07       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.