Literature DB >> 20370851

Intrathoracic impedance preceding ventricular tachyarrhythmia episodes.

Hans J Moore1, Matthew N Peters, Michael R Franz, Pamela E Karasik, Steven N Singh, Ross D Fletcher.   

Abstract

BACKGROUND: Heart failure is associated with ventricular tachyarrhythmias (VT/VF). Fluid accumulation during worsened heart failure may trigger VT/VF. Increased intrathoracic impedance has been correlated with fluid accumulation during heart failure. Implanted defibrillators capable of daily measures of intrathoracic impedance allow correlation of impedance with occurrence of VT/VF. We hypothesized that VT/VF episodes are preceded by decreases in intrathoracic impedance. The goal was to identify the relationship of intrathoracic impedance measured by implanted cardioverter defibrillators to the occurrence of VT/VF.
METHOD: Implanted defibrillator follow-up data were obtained retrospectively. Those with Medtronic OptiVol (Medtronic Inc., Minneapolis, MN, USA), storing averaged daily and reference impedance values, were reviewed for VT/VF episodes. Impedance changes in the week leading up to VT/VF were analyzed.
RESULTS: A total of 317 VT/VF episodes in a cohort of 121 patients' follow-up data were evaluated. Averaged daily intrathoracic impedance declined preceding 64% of VT/VF episodes, with an average decline of 0.46 +/- 0.35 Ohms from the day before the VT/VF episodes. However, the mean values of the averaged daily and reference impedance did not change significantly. A novel measure, DeltaTI, the sum of the daily differences between the averaged daily and reference impedance, was negative preceding 66% of VT/VF episodes (P < 0.001). The mean DeltaTI was -4.0 +/- 1.3 Ohms, which was significantly lower than the theoretically expected value of zero Ohms (P < 0.01).
CONCLUSION: (1) Averaged daily impedance declined preceding 64% of VT/VF episodes, but the overall decline was of small magnitude; (2) a novel measure, DeltaTI, was negative preceding 66% of VT/VF episodes, and significantly below zero.

Entities:  

Mesh:

Year:  2010        PMID: 20370851     DOI: 10.1111/j.1540-8159.2010.02746.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  Congestion is the driving force behind heart failure.

Authors:  Maya Guglin
Journal:  Curr Heart Fail Rep       Date:  2012-09

Review 2.  Heart failure as a substrate and trigger for ventricular tachycardia.

Authors:  Chikezie K Alvarez; Edmond Cronin; William L Baker; Jeffrey Kluger
Journal:  J Interv Card Electrophysiol       Date:  2019-10-09       Impact factor: 1.900

3.  Threshold crossing of device-based intrathoracic impedance trends identifies relatively increased mortality risk.

Authors:  Wai Hong Wilson Tang; Eduardo N Warman; James W Johnson; Roy S Small; James Thomas Heywood
Journal:  Eur Heart J       Date:  2012-06-07       Impact factor: 29.983

4.  Key role of congestion in natural history of heart failure.

Authors:  Maya Guglin
Journal:  Int J Gen Med       Date:  2011-08-15
  4 in total