Literature DB >> 20370641

Cyclic loads do not compromise functionality of the interspinous spacer or cause damage to the spinal segment: an in vitro analysis.

Amanesh Goyal1, Vijay K Goel, Ankit Mehta, David Dick, Suresh Reddy Chinthakunta, Lisa Ferrara.   

Abstract

STUDY
DESIGN: A biomechanical study to evaluate the effects of interspinous spacer under cyclic complex loading.
OBJECTIVES: To determine the risk of device migration and to assess damage on the device and specimen under extreme coupled motion. Another objective was to evaluate the effect on vertebral foramen and canal dimensions after spacer implantation. SUMMARY OF BACKGROUND DATA: Interspinous spacers are a relatively new treatment option that are clinically exposed to complex loads. However, the biomechanical performance of these spacers has not been well characterized.
METHODS: Six human cadaveric motion segments were used for this study. The interspinous spacer (SuperionTM, Vertiflex Inc, California) was tested for 5 degrees extension/10 degrees flexion coupled with an axial rotation of +/-3 degrees. CT images were taken for specimens in neutral, 5 degrees extension, and 10 degrees flexion before and after the implantation of the spacer. Vertebral foramen and canal dimensions were quantified.
RESULTS: There was no device migration or subsidence. Specimens did not sustain any significant injury during testing. Canal area was minimally altered and foramen height, width, and area increased in extension and were statistically significant as compared to intact.
CONCLUSION: Interspinous spacer effectively prevents the motion at the implanted level and does not change the anatomy significantly.

Entities:  

Mesh:

Year:  2008        PMID: 20370641     DOI: 10.1615/jlongtermeffmedimplants.v18.i4.30

Source DB:  PubMed          Journal:  J Long Term Eff Med Implants        ISSN: 1050-6934


  5 in total

1.  Biomechanics of disc degeneration.

Authors:  V Palepu; M Kodigudla; V K Goel
Journal:  Adv Orthop       Date:  2012-06-17

2.  Minimally invasive treatment of lumbar spinal stenosis with a novel interspinous spacer.

Authors:  Shay Shabat; Larry E Miller; Jon E Block; Reuven Gepstein
Journal:  Clin Interv Aging       Date:  2011-09-08       Impact factor: 4.458

3.  Minimally invasive treatment of moderate lumbar spinal stenosis with the superion interspinous spacer.

Authors:  Walter Bini; Larry E Miller; Jon E Block
Journal:  Open Orthop J       Date:  2011-10-14

4.  Interspinous spacer implant in patients with lumbar spinal stenosis: preliminary results of a multicenter, randomized, controlled trial.

Authors:  Larry E Miller; Jon E Block
Journal:  Pain Res Treat       Date:  2012-02-07

Review 5.  A Review of Emerging Evidence for Utilization of a Percutaneous Interspinous Process Decompression Device to Treat Symptomatic Lumbar Adjacent-Segment Degeneration.

Authors:  Timothy R Deer; Dawood Sayed; Mark N Malinowski; Jeffery J Rowe; Jessica B Jameson; Kevin Liang; Joseph A Sclafani
Journal:  Pain Med       Date:  2019-12-01       Impact factor: 3.750

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.