Scanning electron microscopy investigation of PMMA removal by laser irradiation (Er:YAG) in comparison with an ultrasonic system and curettage in hip joint revision arthroplasty.

Klaus Birnbaum1, Norbert Gutknecht.   

Abstract

The cement often left in the femur socket during hip joint revision arthroplasty is usually removed by curettage. Another method for removing the cement is to use an ultrasonic system, and yet another alternative may be to use a laser system. The aim of these investigations was to determine the pulse rate and pulse energy of the Er:YAG laser for sufficient cement ablation. We also compared the results obtained using the laser with those obtained using an ultrasonic device or curettage by histological and scanning electron microscopy (SEM) investigation of the border zone between the polymethyl methacrylate (PMMA) and unfixed specimens of femoral bone. Therefore we prepared 30 unfixed human femur stems after hip joint replacement and prepared ten sagittal sections from each femur stem (in total 300 sections). Of these 300 specimens, 180 were treated with the Er:YAG laser, 60 with the ultrasonic system and 60 by curettage. The high pulse energy of 500 mJ and a pulse rate of 4 Hz provided the highest PMMA ablation rate, although the boundary surface between PMMA and femoral bone was not as fine-grained as found in samples treated at 15 Hz and 250 mJ. However, the treatment time for the same cement ablation rate with the latter settings was twice that at 4 Hz and 500 mJ. Compared to the boundary surfaces treated with the ultrasonic device or curettage, the laser-treated samples had a more distinct undifferentiated boundary surface between PMMA and femoral bone. After development of the Er:YAG-laser to provide higher pulse energies, it may in the future be an additional efficient method for the removal of PMMA in revision arthroplasty. The Er:YAG laser should be combined with an endoscopic and a rinsing suction system so that PMMA can be removed from the femoral shaft under direct vision.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20354888     DOI: 10.1007/s10103-010-0759-z

Source DB:  PubMed          Journal:  Lasers Med Sci        ISSN: 0268-8921            Impact factor:   3.161


  22 in total

1.  Revision arthroplasty facilitated by ultrasonic tool cement removal. Part II: Histologic analysis of endosteal bone after cement removal.

Authors:  J T Caillouette; R S Gorab; R C Klapper; S H Anzel
Journal:  Orthop Rev       Date:  1991-05

2.  Mechanical effects of erbium:YAG laser bone ablation.

Authors:  R Hibst
Journal:  Lasers Surg Med       Date:  1992       Impact factor: 4.025

3.  [Use of ultrasound for the removal of cement in hip prosthesis reoperations].

Authors:  F Honnart
Journal:  Rev Chir Orthop Reparatrice Appar Mot       Date:  1996

4.  Use of high-energy shock waves for bone cement removal.

Authors:  T C May; W R Krause; A J Preslar; M J Smith; A J Beaudoin; J A Cardea
Journal:  J Arthroplasty       Date:  1990-03       Impact factor: 4.757

5.  Er:YAG laser ablation of tissue: effect of pulse duration and tissue type on thermal damage.

Authors:  J T Walsh; T J Flotte; T F Deutsch
Journal:  Lasers Surg Med       Date:  1989       Impact factor: 4.025

6.  Measurements of thermal properties for human femora.

Authors:  S Biyikli; M F Modest; R Tarr
Journal:  J Biomed Mater Res       Date:  1986 Nov-Dec

7.  Ablation of bone and methacrylate by a prototype mid-infrared erbium:YAG laser.

Authors:  J S Nelson; L Yow; L H Liaw; L Macleay; R B Zavar; A Orenstein; W H Wright; J J Andrews; M W Berns
Journal:  Lasers Surg Med       Date:  1988       Impact factor: 4.025

8.  Revision total hip arthroplasty.

Authors:  B F Kavanagh; D M Ilstrup; R H Fitzgerald
Journal:  J Bone Joint Surg Am       Date:  1985-04       Impact factor: 5.284

9.  [Ultrasonic modelling of bone cement - (reoperation of total hip endoprosthesis) (author's transl)].

Authors:  U Weber; M Nietert; E Jacob
Journal:  Z Orthop Ihre Grenzgeb       Date:  1981-06

10.  Carbon dioxide laser removal of polymethylmethacrylate.

Authors:  H H Sherk; G Lane; A Rhodes; J Black
Journal:  Clin Orthop Relat Res       Date:  1995-01       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.