Literature DB >> 20346045

In vivo microwave-induced porcine kidney thermoablation: results and perspectives from a pilot study of a new probe.

Riccardo Bartoletti1, Tommaso Cai, Nevio Tosoratti, Claudio Amabile, Alfonso Crisci, Galliano Tinacci, Nicola Mondaini, Paolo Gontero, Sandro Gelsomino, Gabriella Nesi.   

Abstract

OBJECTIVE: To test the in vivo effects (toxicity, completeness of necrosis, dimensions of the lesion) of microwave thermoablation on porcine kidneys, using the Amica Probe v3 (Hospital Service SpA, Aprilia, Italy), in a refrigerated 17-G microwave applicator, that can be used to induce a spherical necrotic area. PATIENTS AND METHODS: Six pigs were used; each kidney was treated, with no kidney pedicle clamping, by microwave thermoablation at least in three different zones with different exposure times and power, during open surgery. Twelve kidneys had 32 microwave thermoablations overall. The kidneys were then surgically removed, and necrotic lesions measured and evaluated microscopically. The sphericity index (SI) was also calculated to evaluate lesion reproducibility. Areas of renal tissue that were missed were then microscopically evaluated by NADH in vivo staining.
RESULTS: In all, 32 thermoablations were applied; the mean (sd) lesion diameter ranged from 1.2 (0.3) to 4.2 (0.1) cm and changed in relation to both power and time of exposure. The 50-W power particularly induced necrotic renal lesions ranging from 1.9 (0.2) to 4.2 (0.1) cm as a function of the time of exposure and the optimal SI (1.04). Pathological evaluation showed no skipped areas in the context of the lesion, or healthy kidney tissue damage close to necrotic lesions.
CONCLUSIONS: Thermoablation with the Amica probe is safe and showed excellent in vivo effects in this porcine model. Increasing the exposure time at 50 W power could be a useful percutaneous minimally invasive treatment for small solid masses (<4.2 cm), avoiding the risk of missing tumour areas or kidney parenchymal damage from microwave treatment.
© 2010 THE AUTHORS. JOURNAL COMPILATION © 2010 BJU INTERNATIONAL.

Entities:  

Mesh:

Year:  2010        PMID: 20346045     DOI: 10.1111/j.1464-410X.2010.09271.x

Source DB:  PubMed          Journal:  BJU Int        ISSN: 1464-4096            Impact factor:   5.588


  4 in total

Review 1.  Microwave ablation of hepatocellular carcinoma.

Authors:  Guido Poggi; Nevio Tosoratti; Benedetta Montagna; Chiara Picchi
Journal:  World J Hepatol       Date:  2015-11-08

2.  Ten-year experience of percutaneous image-guided radiofrequency ablation of malignant renal tumours in high-risk patients.

Authors:  P Balageas; F Cornelis; Y Le Bras; R Hubrecht; J C Bernhard; J M Ferrière; A Ravaud; N Grenier
Journal:  Eur Radiol       Date:  2013-02-27       Impact factor: 5.315

3.  Realistic anatomical prostate models for surgical skills workshops using ballistic gelatin for nerve-sparing radical prostatectomy and fruit for simple prostatectomy.

Authors:  Nathan Lawrentschuk; Uri Lindner; Laurence Klotz
Journal:  Korean J Urol       Date:  2011-02-21

4.  Efficacy and Safety of Microwave Ablation for Malignant Renal Tumors: An Updated Systematic Review and Meta-Analysis of the Literature Since 2012.

Authors:  Sang Hyun Choi; Jong Woo Kim; Jin Hyoung Kim; Kyung Won Kim
Journal:  Korean J Radiol       Date:  2018-08-06       Impact factor: 3.500

  4 in total