Literature DB >> 20335752

Locked plate fixation of osteoporotic humeral shaft fractures: are two locking screws per segment enough?

David J Hak1, Peter Althausen, Scott J Hazelwood.   

Abstract

OBJECTIVE: The purpose of this study was to compare the biomechanical behavior of using two versus three locking screws per bone segment in a cadaveric humerus fracture gap model.
METHODS: Six matched pairs of elderly osteoporotic fresh-frozen human cadaveric humerii were used. An eight-hole locking compression plate was placed posteriorly on the humeral shaft and secured with either four or six bicortical locking screws. A 5-mm middiaphyseal gap osteotomy was created to simulate a comminuted fracture without bony contact. Specimens were tested in offset axial compression, four-point anteroposterior bending, four-point medial-lateral bending, and torsion. After the initial testing in each of these modalities, the constructs were cyclically loaded in torsion and again tested in the four loading modalities. Lastly, the fixation constructs were then tested to failure in torsion.
RESULTS: There were no significant differences in stiffness between the group fixed with two screws per segment and the group fixed with three screws per segment. The peak torque to failure was higher in the four-screw construct compared with the six-screw construct. The mean torque to failure was 23.5 +/- 3.7 Nm in the construct with two locking screws per segment compared with 20.4 +/- 2.8 Nm in the construct with three locking screws per segment (P = 0.030).
CONCLUSIONS: The addition of a third screw in the locked plate construct did not add to the mechanical stability in axial loading, bending, or torsion. In testing to failure, the addition of a third screw resulted in lower load to failure.

Entities:  

Mesh:

Year:  2010        PMID: 20335752     DOI: 10.1097/BOT.0b013e3181bdd1da

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  12 in total

1.  Antero-lateral minimally invasive plate osteosynthesis (MIPO) with the radial nerve exploration for extra-articular distal-third diaphyseal fractures of the humerus.

Authors:  Wei Zhao; Wei Qu; Chongyang Fu; Huajun Jiang; Sida Liu; Chao Cheng
Journal:  Int Orthop       Date:  2017-06-02       Impact factor: 3.075

2.  [Therapy of humeral shaft fractures].

Authors:  P C Strohm; D C Kubosch; E J Hübner; N P Südkamp; M Jaeger; K Reising
Journal:  Chirurg       Date:  2011-10       Impact factor: 0.955

3.  Dual plating of humeral shaft fractures: orthogonal plates biomechanically outperform side-by-side plates.

Authors:  Victor Kosmopoulos; Arvind D Nana
Journal:  Clin Orthop Relat Res       Date:  2013-11-12       Impact factor: 4.176

4.  Plating of metacarpal fractures with locked or nonlocked screws, a biomechanical study: how many cortices are really necessary?

Authors:  Cameron Barr; Anthony W Behn; Jeffrey Yao
Journal:  Hand (N Y)       Date:  2013-12

5.  Comparison of Screw Quantity and Placement of Metacarpal Fracture Fixation: A Biomechanical Study.

Authors:  Stephen P Canton; Srujan Dadi; Austin Anthony; Ryan T Black; Michael Clancy; John R Fowler
Journal:  Hand (N Y)       Date:  2020-12-21

6.  Single locking compression plate fixation of extra-articular distal humeral fractures.

Authors:  Malhar N Kumar; M R Ravishankar; Ravikiran Manur
Journal:  J Orthop Traumatol       Date:  2014-10-19

7.  Wrist Arthrodesis in Rheumatoid Arthritis Using an LCP Metaphyseal Locking Plate versus an AO Wrist Fusion Plate.

Authors:  Toshitake Taii; Takumi Matsumoto; Sakae Tanaka; Ichiro Nakamura; Katsumi Ito; Takuo Juji
Journal:  Int J Rheumatol       Date:  2018-07-10

8.  Dual plating for fixation of humeral shaft fractures: A mechanical comparison of various combinations of plate lengths.

Authors:  Ahmet Karakasli; Onur Basci; Fatih Ertem; Eyad Sekik; Hasan Havitcioglu
Journal:  Acta Orthop Traumatol Turc       Date:  2016-07-29       Impact factor: 1.511

9.  Fracture fixation with two locking screws versus three non-locking screws: A biomechanical comparison in a normal and an osteoporotic bone model.

Authors:  B Grawe; T Le; S Williamson; A Archdeacon; L Zardiackas
Journal:  Bone Joint Res       Date:  2012-06-01       Impact factor: 5.853

10.  A Biomechanical and Clinical Comparison of Midshaft Clavicle Plate Fixation: Are 2 Screws as Good as 3 on Each Side of the Fracture?

Authors:  Christopher G Larsen; Brian Sleasman; Steven C Chudik
Journal:  Orthop J Sports Med       Date:  2017-09-01
View more