Assessment of gas flow waves for endotracheal tube placement in an ovine model of neonatal resuscitation.

G M Schmölzer1, S B Hooper, K J Crossley, B J Allison, C J Morley, P G Davis.   

Abstract

AIM: Clinical assessment and end-tidal CO(2) (ETCO(2)) detectors are routinely used to verify correct endotracheal tube (ETT) placement. However, ETCO(2) detectors may mislead clinicians by failing to correctly identify placement of an ETT under a variety of circumstances. A flow sensor measures and displays gas flow in and out of an ETT. We compared endotracheal flow sensor recordings with a colorimetric CO(2)-detector (Pedi-Cap) to detect endotracheal intubation in a preterm sheep model of neonatal resuscitation.
METHODS: Six preterm lambs were intubated and ventilated immediately after delivery. At 5 min the oesophagus was also intubated with a similar tube. The endotracheal tube and oesophageal tubes were attached to a Pedi-Cap and flow sensor in random order. Two observers, blinded to the positions of the tubes, used a ETCO(2) detector and the flow sensor recording to determine whether the tube was in the trachea or oesophagus. The experiment was repeated 10 times for each animal. In the last three animals (30 recordings) the number of inflations required to correctly identify the tube placement was noted.
RESULTS: The Pedi-Cap and the flow sensor correctly identified tube placement in all studies. Thus, the sensitivity, specificity, and positive and negative predictive values of both devices were 100%. At least three, and up to 10, inflations were required to identify tube location with the Pedi-Cap compared to one or two inflations with the flow sensor.
CONCLUSION: A flow sensor correctly identifies tube placement within the first two inflations. The Pedi-Cap required more inflations to correctly identify tube placement. Copyright 2010 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20334964     DOI: 10.1016/j.resuscitation.2010.02.018

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  4 in total

1.  A novel airway device with tactile sensing capabilities for verifying correct endotracheal tube placement.

Authors:  Pauwel Goethals; Harshu Chaobal; Dominiek Reynaerts; David Schaner
Journal:  J Clin Monit Comput       Date:  2013-09-25       Impact factor: 2.502

2.  Assessment of endotracheal tube placement in newborn infants: a randomized controlled trial.

Authors:  S van Os; P-Y Cheung; K Kushniruk; M O'Reilly; K Aziz; G M Schmölzer
Journal:  J Perinatol       Date:  2016-01-14       Impact factor: 2.521

Review 3.  WITHDRAWN: Techniques to ascertain correct endotracheal tube placement in neonates.

Authors:  Georg M Schmölzer; Charles Christoph C Roehr
Journal:  Cochrane Database Syst Rev       Date:  2018-07-05

4.  Noninvasive Monitoring during Interhospital Transport of Newborn Infants.

Authors:  Georg M Schmölzer; Megan O'Reilly; Po-Yin Cheung
Journal:  Crit Care Res Pract       Date:  2013-02-24