H Ashraf1, B de Hoop, S B Shaker, A Dirksen, K S Bach, H Hansen, M Prokop, J H Pedersen.
Abstract
OBJECTIVE: We examined the reproducibility of lung nodule volumetry software that offers three different volumetry algorithms.Entities:
Mesh:
Year: 2010 PMID: 20306082 PMCID: PMC2899012 DOI: 10.1007/s00330-010-1749-z
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Screenshot of Siemens LungCARE software. The right lower window displays a visual 3D presentation of the nodule
Algorithm applied by two readers, nodule characteristics and differences between readings
| Algorithm of 1st reader | Algorithm of 2nd reader | No. of nodule readings, (%) | Volume of nodules, Mean (range) mm3 | Difference between readers (1st – 2nd) | ||||
|---|---|---|---|---|---|---|---|---|
| Mean (SD) | Range | No difference in volume | Difference in volume >25% | |||||
| Percent of minimal reading | ||||||||
| Same algorithms | Small size | Small size | 252 (65) | 160 (7–2,088) | −1 (11) | −65 to 48 | 58% | 4% |
| All sizes | All sizes | 36 (9) | 1,056 (146–3,486) | 1 (6) | −14 to 31 | 22% | 3% | |
| Subsolid | Subsolid | 23 (6) | 392 (51–1,296) | −1 (3) | −5 to 8 | 4% | 0% | |
| Total | 311 (80) | 281 (7–3,486) | 0 (10) | −65 to 48 | 50% | 4% | ||
| Different algorithms | Small size | All sizes | 35 (9) | 255 (6–995) | −100 (154) | −890 to −6 | 0% | 77% |
| All sizes | Small size | 34(8) | 304 (9–1,780) | 77 (74) | 14–406 | 0% | 82% | |
| Small/All sizes | Subsolid | 11 (3) | 249 (5–2,107) | −1,428 (1,368) | −4,379 to −196 | 0% | 100% | |
| Subsolid | Small/All sizes | 0 (0) | – | – | – | – | – | |
| Total | 80 (20) | 275 (5-2,107) | −207 (705) | −4,379 to 406 | 0% | 83% | ||
Fig. 2Bland-Altman plots when the same algorithm was chosen by both readers. Small size = 252 nodules, All sizes = 36 nodules and Subsolid = 23 nodules. Dotted lines indicate 25% variation. Because of the relationship between nodule size and variability (and to simplify), lines corresponding to the 95% confidence interval (CI) were omitted
Fig. 3Bland-Altman plots when the readers chose different algorithms. Small size/All sizes = 35 nodules, All sizes/Small size = 34 nodules and All sizes/Small size – Subsolid = 11 nodules. Dotted lines indicate 25% variation. Because of the relationship between nodule size and variability (and to simplify), lines corresponding to the 95% CI were omitted
Fig. 4The same nodule analysed at the same time with the Subsolid (a), Small size (b) and All sizes (c) algorithms