Microemboli, developed during haemodialysis, pass the lung barrier and may cause ischaemic lesions in organs such as the brain.

Ulf Forsberg1, Per Jonsson, Christofer Stegmayr, Bernd Stegmayr.   

Abstract

BACKGROUND: Chronic haemodialysis (HD) may relieve some medical problems of terminal uraemia, but the life expectancy of patients is still significantly shortened, and there is a greatly increased morbidity. This includes pulmonary morbidity and chronic central nervous system (CNS) abnormalities. Previous studies have shown that a considerable amount of air microbubbles emanate within the blood lines of the dialysis device and pass the air detector without sounding an alarm. The aim of this study was to investigate whether microemboli can pass to the patient and whether they could be detected in the carotid artery.
METHODS: A total of 54 patients on chronic HD (16 with central dialysis catheter) were investigated with an ultrasound detector (Hatteland, Røyken, Norway) for the presence of microemboli at the arteriovenous (AV) fistula/graft and at the common carotid artery before and during HD. Measurements were taken for 2 and 5 min, respectively. Non-parametric paired statistics were used (Wilcoxon).
RESULTS: The median number (range) and mean +/- SD of microembolic signals detected at the AV access site before commencing dialysis and during HD were 0 (0-3) and 0.2+/- 0.5 versus 4 (0-85) and 13.5 +/- 20 (P = 0.000); at the carotid artery, 1 (0-14) and 1.7 +/- 2.9 versus 2 (0-36) and 3.5 +/- 5.8 (P = 0.008).
CONCLUSIONS: The infused and returning fluid from HD devices contains air microbubbles that enter the patient without triggering any alarms. These small emboli pass the lung and may cause ischaemic lesions in organs supported by the arterial circuit, such as the brain.

Entities:  

Mesh:

Year:  2010        PMID: 20305135     DOI: 10.1093/ndt/gfq116

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  6 in total

1.  Factors associated with favorable response to hyperbaric oxygen therapy among patients presenting with iatrogenic cerebral arterial gas embolism.

Authors:  Wondwossen G Tekle; Cheryl D Adkinson; Saqib A Chaudhry; Vikram Jadhav; Ameer E Hassan; Gustavo J Rodriguez; Adnan I Qureshi
Journal:  Neurocrit Care       Date:  2013-04       Impact factor: 3.210

2.  Cerebral microembolism in the critically ill with acute kidney injury (COMET-AKI trial): study protocol for a randomized controlled clinical trial.

Authors:  Gabor Erdoes; Dominik E Uehlinger; Beatrice Kobel; Monika P Stucki; Roland Wiest; Frank Stueber; Niklaus Fankhauser; Stephan M Jakob; Joerg C Schefold
Journal:  Trials       Date:  2018-03-21       Impact factor: 2.279

3.  Cortical Blindness and Altered Mental Status following Routine Hemodialysis, a Case of Iatrogenic Cerebral Air Embolism.

Authors:  Lawrence Lau; Kory London
Journal:  Case Rep Emerg Med       Date:  2018-03-14

4.  Observation of microbubbles during standard dialysis treatments.

Authors:  Stephan Wagner; Christiane Rode; Ralf Wojke; Bernard Canaud
Journal:  Clin Kidney J       Date:  2015-07-03

5.  Evaluation of low-volume post-dilution online hemodiafiltration with Japanese high-performance hemodiafilters.

Authors:  Kenji Sakurai; Hiromi Hosoya; Yoshitaka Kurihara; Fumi Yamauchi; Ayumi Suzuki; Kaori Kurosawa; Takeshi Saito
Journal:  J Artif Organs       Date:  2020-04-07       Impact factor: 1.731

6.  Interdialytic weight gain of less than 2.5% seems to limit cardiac damage during hemodialysis.

Authors:  Junko Goto; Ulf Forsberg; Per Jonsson; Kenichi Matsuda; Bo Nilsson; Kristina Nilsson Ekdahl; Michael Y Henein; Bernd G Stegmayr
Journal:  Int J Artif Organs       Date:  2020-12-18       Impact factor: 1.595

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.