| Literature DB >> 20300736 |
Ibolya Kiss1, Ivett Bacskay, Ferenc Kilár, Attila Felinger.
Abstract
The characterization of mass-transfer processes in a chromatographic column during a separation process is essential, since the influence of the mass-transfer kinetics on the shape of the chromatographic band profiles and on the efficiency of the separation is crucial. Several sources of mass transfer in a chromatographic bed have been identified and studied: the axial dispersion in the stream of mobile phase, the external mass-transfer resistance, intraparticle diffusion, and the kinetics of adsorption-desorption. We measured and compared the characteristics and performance of a new brand of shell particles and those of a conventional brand of totally porous silica particles. The shell stationary phase was made of 2.7-microm superficially porous particles (a 1.7-microm solid core is covered with a 0.5-microm-thick shell of porous silica). The other material consisted of totally porous particles of conventional 3.5-microm commercial silica. We measured the first and second central moments of the peaks of human insulin over a wide range of mobile phase velocities (from 0.02 to 1.3 mL/min) at 20 degrees C. The plate height equations were constructed and the axial dispersion, external mass transfer, as well as the intraparticle diffusion coefficients were calculated for the two stationary phases.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20300736 DOI: 10.1007/s00216-010-3627-6