| Literature DB >> 2026715 |
G Durand1, N de Bertrand, D Barceló.
Abstract
The effect of four different mobile phase compositions with reversed-phase methanol-water (50:50) + 0.05 M ammonium acetate, methanol-water (50:50) + 0.05 M ammonium formate, acetonitrile-water (50:50) + 0.05 M ammonium acetate and acetonitrile-water (50:50) + 0.05 M ammonium formate were compared in filament-on thermospray liquid chromatography-mass spectrometry for the determination of carbamate and chlorotriazine pesticides. In the positive-ion mode, [M + H]+ and [M + NH4]+ were generally the base peaks for the chlorotriazines and the carbamates, respectively. Depending on the mobile phase used, other adduct ions obtained corresponded to [M + CH3CN + H]+, [M + CH3OH + NH4]+, [M + CH3COONH4 + NH4 - 2H2O]+, [M + CH3CN + NH4]+, [M + CH3COONH4 + H - H2O]+ and the dimer [2M + H]+. In the negative-ion mode, [M - H]- and adducts with the ionizing additive [M + CH3COO]- or [M + HCOO]- were obtained. Other ions for the carbamates carbaryl and oxamyl corresponded to [M - CONHCH3 + CH3COOH]- and [M - CON(CH3)2 + HCOO]-, respectively. The variation of mobile phase composition provides additional structural information in thermospray liquid chromatography-mass spectrometry with no appreciable loss of sensitivity. Applications are reported for the determination of carbamate and chlorotriazine pesticides at the ng/g level in spiked and real soil samples, respectively.Entities:
Mesh:
Substances:
Year: 1991 PMID: 2026715 DOI: 10.1016/0378-4347(91)80603-a
Source DB: PubMed Journal: J Chromatogr