Literature DB >> 20234083

Atomic force microscopy study of immunosensor surface to scale down the size of ELISA-type sensors.

D Volkov1, G Strack, J Halámek, E Katz, I Sokolov.   

Abstract

Here we describe the use of atomic force microscopy (AFM) to study the nanoscale mechanics of the molecular layers of a popular immunosensor, ELISA (enzyme-linked immunosorbent assay) type. We characterize the sensor surface in terms of brush length and grafting density of the molecular layers. The obtained data demonstrated that a reliable reading of the immunosignal (a suggested dimensionless combination of brush length and grafting density) can be attained from an area as small as approximately 3 microm(2). This is approximately 4 million times smaller compared to typical ELISA sensors. The immunosensor described is composed of a molecular mix of two different antigens. Intriguingly, we find that AFM can reliably distinguish between having the immunosignal from either antibody and from both antibodies together. This was impossible to get by using standard optical detection methods.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20234083     DOI: 10.1088/0957-4484/21/14/145503

Source DB:  PubMed          Journal:  Nanotechnology        ISSN: 0957-4484            Impact factor:   3.874


  1 in total

1.  A total internal reflection ellipsometry and atomic force microscopy study of interactions between Proteus mirabilis lipopolysaccharides and antibodies.

Authors:  J Gleńska-Olender; S Sęk; K Dworecki; W Kaca
Journal:  Eur Biophys J       Date:  2015-04-09       Impact factor: 1.733

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.