Literature DB >> 20233266

Reliability and reproducibility of QRS duration in the selection of candidates for cardiac resynchronization therapy.

Maxime De Guillebon1, Jean-Benoit Thambo, Sylvain Ploux, Antoine Deplagne, Frederic Sacher, Pierre Jais, Michel Haissaguerre, Philippe Ritter, Jacques Clementy, Pierre Bordachar.   

Abstract

BACKGROUND: A QRS >120 ms remains the recommended criterion for the selection of cardiac resynchronization therapy (CRT) candidates. However, the reproducibility of this measurement has not been studied thoroughly.
METHODS: QRS duration was measured by 3 experienced cardiologists and by automatic measurement on 228 electrocardiograms (ECGs) randomly collected from 188 subjects, including neonates, healthy adults, patients with complete and incomplete bundle branch block, and CRT candidates. All ECGs were recorded at a 25 mm/s sweep speed. Forty recordings were duplicated and 50 ECGs were recorded at both 25 and 50 mm/s.
RESULTS: Significant interobserver differences (P < 0.001) were found between each combination of paired observers, with an up to 50-ms absolute variability between cardiologists and low concordance with computerized measurements. Intraobserver absolute variability was also significant (P < 0.01) for the 3 observers. These significant differences persisted (P < 0.01) when focusing our interest on the ECGs in the 100-140 ms range (defined as at least one out of the 4 measures in this range). Considering the 120 ms limit, 22 (27.5%) ECGs were differently classified by at least one of the cardiologists. We observed similar interobserver differences between each combination of paired observers with a 50 mm/s sweep speed.
CONCLUSION: Manual QRS duration measurements were associated with significant inter- and intraobserver variability and low concordance with computerized measurements. The measurement of QRS is, therefore, operator-dependent and a reevaluation of the measurement methods may be essential to develop clinical and investigative standards.

Entities:  

Mesh:

Year:  2010        PMID: 20233266     DOI: 10.1111/j.1540-8167.2010.01743.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  12 in total

Review 1.  Cardiac resynchronization therapy: the issue of non-response.

Authors:  Luigi Padeletti; Alessandro Paoletti Perini; Edoardo Gronda
Journal:  Heart Fail Rev       Date:  2012-01       Impact factor: 4.214

2.  Multiple serial ECGs aid with the diagnosis and prognosis of Brugada syndrome.

Authors:  Jesus Castro Hevia; Margarita Dorantes Sanchez; Frank Martinez Lopez; Osmin Castañeda Chirino; Roylan Falcon Rodriguez; Marcelo Puga Bravo; Joanna de Zayas Galguera; Charles Antzelevitch
Journal:  Int J Cardiol       Date:  2018-08-30       Impact factor: 4.164

Review 3.  Electrical manipulation of the failing heart.

Authors:  Valerio Zacà; Theodore Murphy; Mauro Biffi
Journal:  Heart Fail Rev       Date:  2018-11       Impact factor: 4.214

4.  Different Methods to Measure QRS Duration in CRT Patients: Impact on the Predictive Value of QRS Duration Parameters.

Authors:  Jan De Pooter; Milad El Haddad; Liesbeth Timmers; Frédéric Van Heuverswyn; Luc Jordaens; Mattias Duytschaever; Roland Stroobandt
Journal:  Ann Noninvasive Electrocardiol       Date:  2015-09-22       Impact factor: 1.468

5.  QRS Complex Detection and Measurement Algorithms for Multichannel ECGs in Cardiac Resynchronization Therapy Patients.

Authors:  Antonia E Curtin; Kevin V Burns; Alan J Bank; Theoden I Netoff
Journal:  IEEE J Transl Eng Health Med       Date:  2018-06-05       Impact factor: 3.316

6.  Precision of automated QRS duration measurement in patients treated with cardiac resynchronization therapy.

Authors:  Casper Lund-Andersen; Helen H Petersen; Christian Jøns; Berit T Philbert; Jacob Tfelt-Hansen; Lene T Skovgaard; Jesper H Svendsen
Journal:  J Interv Card Electrophysiol       Date:  2018-03-05       Impact factor: 1.900

7.  Reproducibility of measuring QRS duration and implications for optimization of interventricular pacing delay in cardiac resynchronization therapy.

Authors:  Charlotte Stephansen; Mads Brix Kronborg; Christoffer Tobias Witt; Jens Kristensen; Christian Gerdes; Anders Sommer; Jesper Møller Jensen; Jens Cosedis Nielsen
Journal:  Ann Noninvasive Electrocardiol       Date:  2018-12-06       Impact factor: 1.468

8.  Narrowing filtered QRS duration on signal-averaged electrocardiogram predicts outcomes in cardiac resynchronization therapy patients with nonischemic heart failure.

Authors:  Atsushi Suzuki; Tsuyoshi Shiga; Daigo Yagishita; Yoshimi Yagishita-Tagawa; Kotaro Arai; Yuji Iwanami; Koichiro Ejima; Kyomi Ashihara; Morio Shoda; Nobuhisa Hagiwara
Journal:  Ann Noninvasive Electrocardiol       Date:  2017-11-30       Impact factor: 1.468

9.  Statistical evaluation of reproducibility of automated ECG measurements: an example from arrhythmogenic right ventricular dysplasia/cardiomyopathy clinic.

Authors:  Timothy Huang; Cynthia A James; Crystal Tichnell; Brittney Murray; Joel Xue; Hugh Calkins; Larisa G Tereshchenko
Journal:  Biomed Signal Process Control       Date:  2014-09-01       Impact factor: 3.880

10.  A new electrocardiogram marker to identify patients at low risk for ventricular tachyarrhythmias: sum magnitude of the absolute QRST integral.

Authors:  Larisa G Tereshchenko; Alan Cheng; Barry J Fetics; Barbara Butcher; Joseph E Marine; David D Spragg; Sunil Sinha; Darshan Dalal; Hugh Calkins; Gordon F Tomaselli; Ronald D Berger
Journal:  J Electrocardiol       Date:  2010-11-20       Impact factor: 1.438

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.