A photon spectrometric dose-rate constant determination for the Advantage Pd-103 brachytherapy source.

Zhe Jay Chen1, Paul Bongiorni, Ravinder Nath.   

Abstract

PURPOSE: Although several dosimetric characterizations using Monte Carlo simulation and thermoluminescent dosimetry (TLD) have been reported for the new Advantage Pd-103 source (IsoAid, LLC, Port Richey, FL), no AAPM consensus value has been established for the dosimetric parameters of the source. The aim of this work was to perform an additional dose-rate constant (lamda) determination using a recently established photon spectrometry technique (PST) that is independent of the published TLD and Monte Carlo techniques.
METHODS: Three Model IAPD-103A Advantage Pd-103 sources were used in this study. The relative photon energy spectrum emitted by each source along the transverse axis was measured using a high-resolution germanium spectrometer designed for low-energy photons. For each source, the dose-rate constant was determined from its emitted energy spectrum. The PST-determined dose-rate constant (PST lamda) was then compared to those determined by TLD (TLD lamda) and Monte Carlo (MC lamda) techniques. A likely consensus lamda value was estimated as the arithmetic mean of the average lamda values determined by each of three different techniques.
RESULTS: The average PST lamda value for the three Advantage sources was found to be (0.676 +.- 0.026) cGyh(-1) U(-1). Intersource variation in PST lamda was less than 0.01%. The PST lamda was within 2% of the reported MC lamda values determined by PTRAN, EGSnrc, and MCNP5 codes. It was 3.4% lower than the reported TLD lamda. A likely consensus lamda value was estimated to be (0.688 +/- 0.026) cGyh(-1) U(-1), similar to the AAPM consensus values recommended currently for the Theragenics (Buford, GA) Model 200 (0.686 +/- 0.033) cGyh(-1) U(-1), the NASI (Chatsworth, CA) Model MED3633 (0.688 +/- 0.033) cGyh(-1) U(-1), and the Best Medical (Springfield, VA) Model 2335 (0.685 +/- 0.033) cGyh(-1) U(-1) 103Pd sources.
CONCLUSIONS: An independent lamda determination has been performed for the Advantage Pd-103 source. The PST lamda obtained in this work provides additional information needed for establishing a more accurate consensus lamda value for the Advantage Pd-103 source.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20229876      PMCID: PMC2821420          DOI: 10.1118/1.3298011

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations.

Authors:  Mark J Rivard; Bert M Coursey; Larry A DeWerd; William F Hanson; M Saiful Huq; Geoffrey S Ibbott; Michael G Mitch; Ravinder Nath; Jeffrey F Williamson
Journal:  Med Phys       Date:  2004-03       Impact factor: 4.071

2.  Supplement to the 2004 update of the AAPM Task Group No. 43 Report.

Authors:  Mark J Rivard; Wayne M Butler; Larry A DeWerd; M Saiful Huq; Geoffrey S Ibbott; Ali S Meigooni; Christopher S Melhus; Michael G Mitch; Ravinder Nath; Jeffrey F Williamson
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

3.  Photon spectrometry for the determination of the dose-rate constant of low-energy photon-emitting brachytherapy sources.

Authors:  Zhe Jay Chen; Ravinder Nath
Journal:  Med Phys       Date:  2007-04       Impact factor: 4.071

4.  An EGSnrc Monte Carlo-calculated database of TG-43 parameters.

Authors:  R E P Taylor; D W O Rogers
Journal:  Med Phys       Date:  2008-09       Impact factor: 4.071

5.  Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine.

Authors:  R Nath; L L Anderson; G Luxton; K A Weaver; J F Williamson; A S Meigooni
Journal:  Med Phys       Date:  1995-02       Impact factor: 4.071

6.  Theoretical and experimental determination of dosimetric characteristics for ADVANTAGE Pd-103 brachytherapy source.

Authors:  Ali S Meigooni; Sharifeh A Dini; Shahid B Awan; Kai Dou; Rafiq A Koona
Journal:  Appl Radiat Isot       Date:  2006-05-18       Impact factor: 1.513

7.  Monte Carlo dosimetric characterization of the IsoAid ADVANTAGE 103Pd brachytherapy source.

Authors:  Keith T Sowards
Journal:  J Appl Clin Med Phys       Date:  2007-03-20       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.