Catheter ablation of atrioventricular junction using radiofrequency current in 17 patients. Comparison of standard and large-tip catheter electrodes.

W M Jackman1, X Z Wang, K J Friday, D M Fitzgerald, C Roman, K Moulton, P D Margolis, A J Bowman, K H Kuck, G V Naccarelli.   

Abstract

BACKGROUND: Two catheter electrode systems were compared for delivering radiofrequency current for ablation of the atrioventricular junction. Seventeen patients with drug-resistant supraventricular tachyarrhythmias were studied. METHODS AND
RESULTS: A 6F or 7F catheter with six or eight standard electrodes (1.25 mm wide, 2.5-mm spacing) was used in the first seven patients (group 1). A 7F quadripolar catheter with a large-tip electrode (4 mm long; surface area, 27 mm2) was used in the final 10 patients (group 2). Both ablation catheters were positioned to record a large atrial potential and a small but sharp His bundle potential from the distal bipolar electrode pair. Radiofrequency current was applied between a large skin electrode on the left posterior chest and either 1) each individual electrode on the standard-tip electrode catheter at 40 V (group 1) or 2) the large-tip electrode at 50-60 V (group 2). Radiofrequency current was limited to 40 V in group patients because of the strong potential for an early impedance rise when higher voltage is applied through standard electrodes. Complete atrioventricular block was achieved in six of seven group 1 patients and all 10 group 2 patients. A junctional escape rhythm followed ablation in five or six group 1 patients (mean cycle length, 1,066 +/- 162 msec) and eight of 10 group 2 patients (mean cycle length, 1,281 +/- 231 msec). Atrioventricular block was produced in a mean of 4.7 +/- 4.6 radiofrequency current applications delivered over a period of 42 +/- 45 minutes using the large-tip electrode (group 2) compared with 46 +/- 22 applications using standard electrodes (15.9 +/- 10.2 applications delivered through the standard-tip electrode) over a period of 147 +/- 59 minutes (group 1). For the application producing atrioventricular block, the large-tip electrode used higher voltage (58 +/- 17 versus 38 +/- 5 V, p less than 0.03) and had lower impedance (103 +/- 22 versus 148 +/- 40 omega, p less than 0.01), resulting in greater power (33.0 +/- 13.0 versus 10.2 +/- 0.6 W, p less than 0.003) and shorter time to block (8 +/- 3 versus 22 +/- 3 seconds, p less than 0.001). Current delivery through standard electrodes was limited by an impedance rise occurring 7 +/- 7 seconds after the onset of one or more radiofrequency current applications at 10 +/- 1 W in six of seven patients. Using the large-tip electrode, an impedance rise occurred in five of 10 patients, but at 25 +/- 10 W and after 21 +/- 9 seconds. Atrioventricular block occurred before the impedance rise in three of these five patients. Complete atrioventricular block persisted in 15 of 16 patients at a mean follow-up of 8.7 months. Atrioventricular conduction returned at 1 month in one group 2 patient and was successfully ablated by a second procedure. Three group 1 patients died 0.5-2 months after ablation, and a fourth patient underwent cardiac transplantation after 10 months. Pathological examination of the heart in two of these patients showed necrosis of the atrioventricular node and origin of the His bundle, without injury to the middle or distal His bundle. All 10 group 2 patients are alive and subjectively improved after ablation.
CONCLUSIONS: We conclude that catheter-delivered radiofrequency current effectively produces complete atrioventricular block (94%) without requiring general anesthesia or the risk of ventricular dysfunction or cardiac perforation. The large-tip electrode allows a threefold increase in delivered power and markedly decreases the number of pulses and time required to produce atrioventricular block.

Entities:  

Mesh:

Year:  1991        PMID: 2022016     DOI: 10.1161/01.cir.83.5.1562

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  14 in total

Review 1.  Unipolar recording in cardiac electrophysiologic studies.

Authors:  F M Kusumoto
Journal:  J Interv Card Electrophysiol       Date:  1999-07       Impact factor: 1.900

2.  Parahisian radiofrequency catheter ablation in dogs: comparison of the above-valve and below-valve approaches.

Authors:  R Kato; K Matsumoto; O Goktekin; H Matsuo; H Watanabe; M Takahama
Journal:  J Interv Card Electrophysiol       Date:  2000-06       Impact factor: 1.900

3.  A comparative study of radiofrequency ablation in unipolar and bipolar fashion.

Authors:  Y G Wang; Z Y Lu; H Y Zhao; Y E Song; R L Li
Journal:  J Tongji Med Univ       Date:  1995

Review 4.  Non-pharmacological, non-ablative approaches for the treatment of atrial fibrillation: experimental evidence and potential clinical implications.

Authors:  Benjamin J Scherlag; Hiroshi Nakagawa; Warren M Jackman; Ralph Lazzara; Sunny S Po
Journal:  J Cardiovasc Transl Res       Date:  2010-11-06       Impact factor: 4.132

Review 5.  [Catheter ablation of cardiac arrhythmias : Forms of energy and biophysical principles].

Authors:  Thomas Beiert; Jan W Schrickel
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2019-11-06

Review 6.  Maintaining stability of sinus rhythm in atrial fibrillation: antiarrhythmic drugs versus ablation.

Authors:  Gerald V Naccarelli; John Hynes; Deborah L Wolbrette; Luna Bhatta; Mazhar Khan; Jerry Luck
Journal:  Curr Cardiol Rep       Date:  2002-09       Impact factor: 2.931

7.  Prospective study of left ventricular function after radiofrequency ablation of atrioventricular junction in patients with atrial fibrillation.

Authors:  M Edner; K Caidahl; L Bergfeldt; B Darpö; N Edvardsson; M Rosenqvist
Journal:  Br Heart J       Date:  1995-09

Review 8.  Clinical results with catheter ablation: AV junction, atrial fibrillation and ventricular tachycardia.

Authors:  Jonathan Weinstock; Paul J Wang; Munther K Homoud; Mark S Link; N A Mark Estes
Journal:  J Interv Card Electrophysiol       Date:  2003-10       Impact factor: 1.900

9.  The Ablate and Pace Trial: a prospective study of catheter ablation of the AV conduction system and permanent pacemaker implantation for treatment of atrial fibrillation. APT Investigators.

Authors:  G N Kay; K A Ellenbogen; M Giudici; M M Redfield; L S Jenkins; M Mianulli; B Wilkoff
Journal:  J Interv Card Electrophysiol       Date:  1998-06       Impact factor: 1.900

10.  Effect of catheter tip length and position on lesion volume in temperature controlled RF ablation in canine tricuspid valve annulus.

Authors:  A Mehdirad; J Gaiser; P Baker; S West; L Lehmkuhl; P Yong; J Meimer; S Nelson
Journal:  J Interv Card Electrophysiol       Date:  1998-09       Impact factor: 1.900

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.