Literature DB >> 20219581

A biomechanical analysis of titanium miniplates used for treatment of mandibular symphyseal fractures with the finite element method.

Baohui Ji1, Chun Wang, Lei Liu, Jie Long, Weidong Tian, Hang Wang.   

Abstract

OBJECTIVE: This study aimed to evaluate the stress distribution and stress shielding effect of titanium miniplates used for the treatment of symphyseal fractures using finite element (FE) analysis. STUDY
DESIGN: Two 3-D FE models of symphyseal fractured mandibles reduced by technique 1, reduction with a single miniplate, and technique 2, reduction with 2 miniplates, respectively, were developed. Three basic loading conditions were simulated.
RESULTS: The ratios of stress shielding of miniplates were different. Ratios of the lower miniplates in technique 2 were much higher than the upper miniplates and the miniplates in technique 1 during all conditions, and that value of the lower miniplate gained a maximum value of 83.34% during left unilateral molar clenching. The stress areas were concentrated on the central section of the miniplates. However, the stress distribution varied with masticatory conditions.
CONCLUSION: The study demonstrated that miniplate stress distribution and stress shielding effect ratio were affected not only by the way in which the mandible was loaded but also by the number of the miniplates fixing the fracture. Copyright 2010 Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20219581     DOI: 10.1016/j.tripleo.2009.11.003

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  4 in total

1.  Role of 1.5 mm microplates in treatment of symphyseal fracture of mandible: A stress analysis based comparative study.

Authors:  Syed S Ahmed; Siddharth Bhardwaj; Md Kalim Ansari; Omar Farooq; Abid Ali Khan
Journal:  J Oral Biol Craniofac Res       Date:  2017-04-14

Review 2.  Bioabsorbable osteofixation for orthognathic surgery.

Authors:  Young-Wook Park
Journal:  Maxillofac Plast Reconstr Surg       Date:  2015-02-19

3.  Assessment of Lingual Stability in Mandible Fracture: Monocortical Versus Bicortical Fixation Using FEM Analysis.

Authors:  Udupikrishna Joshi; Manju Kurakar
Journal:  J Maxillofac Oral Surg       Date:  2017-12-16

4.  Finite element analysis of locking plate and 1/4 tubular plate for first tarsometatarsal joint fracture-dislocation.

Authors:  Xiao Yu; Wei-Long Li; Qing-Jiang Pang; Rong-Li Zhou
Journal:  J Int Med Res       Date:  2017-07-31       Impact factor: 1.671

  4 in total