Literature DB >> 20219201

Coupling of hydrodynamically closed large bore capillary isotachophoresis with electrospray mass spectrometry.

Roman Tomás1, Marian Koval, Frantisek Foret.   

Abstract

Capillary isotachophoresis with coupled columns provides efficient means for rapid electrophoretic analysis of sample volumes of up to 10microl or more. Commercially available instruments are commonly equipped with conductivity and UV absorbance detectors; however, their on-line coupling with electrospray mass spectrometry is highly desirable. In this work we have modified the commercial coupled column isotachophoresis system for direct connection to an ion trap mass spectrometer. The design included attachment of an elution block with a short capillary transfer line directing the separated zones towards the mass spectrometer. The modification further included separation of the injection and electrode blocks from the separation columns by semipermeable membranes eliminating unwanted fluid movements in the wide bore fluoropolymer separation capillaries. Fused silica capillaries with varying internal diameter were connected as a transfer line between the elution block and mass spectrometer. The transfer line served also as the ESI tip of the sheathless electrospray interface. During the analysis the first, wide bore preseparation capillary with 0.8mm internal diameter served for removal of the bulk sample components and preseparation of the potentially interfering analytes. After the electronic column switching the separation was finished in a 0.3mm internal diameter capillary and the separated ITP zones were transferred in-line by a spray liquid towards the mass spectrometer. The instrumentation was tested for determination of vitamins in whole blood analysis and separation of tryptic peptides. Copyright (c) 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20219201     DOI: 10.1016/j.chroma.2010.02.021

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  5 in total

1.  Preconcentration and detection of the phosphorylated forms of cardiac troponin I in a cascade microchip by cationic isotachophoresis.

Authors:  Danny Bottenus; Mohammad Robiul Hossan; Yexin Ouyang; Wen-Ji Dong; Prashanta Dutta; Cornelius F Ivory
Journal:  Lab Chip       Date:  2011-09-21       Impact factor: 6.799

Review 2.  Recent advances in coupling capillary electrophoresis-based separation techniques to ESI and MALDI-MS.

Authors:  Xuefei Zhong; Zichuan Zhang; Shan Jiang; Lingjun Li
Journal:  Electrophoresis       Date:  2013-12-05       Impact factor: 3.535

3.  10,000-fold concentration increase of the biomarker cardiac troponin I in a reducing union microfluidic chip using cationic isotachophoresis.

Authors:  Danny Bottenus; Talukder Zaki Jubery; Yexin Ouyang; Wen-Ji Dong; Prashanta Dutta; Cornelius F Ivory
Journal:  Lab Chip       Date:  2011-03-07       Impact factor: 6.799

4.  10,000-fold concentration increase in proteins in a cascade microchip using anionic ITP by a 3-D numerical simulation with experimental results.

Authors:  Danny Bottenus; Talukder Zaki Jubery; Prashanta Dutta; Cornelius F Ivory
Journal:  Electrophoresis       Date:  2011-02-10       Impact factor: 3.535

5.  Cationic isotachophoresis separation of the biomarker cardiac troponin I from a high-abundance contaminant, serum albumin.

Authors:  Thomas Jacroux; Danny Bottenus; Bennett Rieck; Cornelius F Ivory; Wen-Ji Dong
Journal:  Electrophoresis       Date:  2014-06-05       Impact factor: 3.535

  5 in total