Mechanical performance comparison between RotaFlow and CentriMag centrifugal blood pumps in an adult ECLS model.

Robert McCoach, Allen Kunselman, Aly El-Banayosy, Akif Undar.   

Abstract

INTRODUCTION: Centrifugal blood pumps have been widely adopted in conventional adult cardiopulmonary bypass and circulatory assist procedures. Different brands of centrifugal blood pumps incorporate distinct designs which affect pump performance. In this adult extracorporeal life support (ECLS) model, the performances of two brands of centrifugal blood pump (RotaFlow blood pump and CentriMag blood pump) were compared.
METHODS: The simulated adult ECLS circuit used in this study included a centrifugal blood pump, Quadrox D membrane oxygenator and Sorin adult ECLS tubing package. A Sorin Cardiovascular(R) VVR(R) 4000i venous reservoir (Sorin S.p.A., Milan, Italy) with a Hoffman clamp served as a pseudo-patient. The circuit was primed with 900ml heparinized human packed red blood cells and 300ml lactated Ringer's solution (total volume 1200 ml, corrected hematocrit 40%). Trials were conducted at normothermia (36 degrees C). Performance, including circuit pressure and flow rate, was measured for every setting analyzed.
RESULTS: The shut-off pressure of the RotaFlow was higher than the CentriMag at all measurement points given the same rotation speed (p < 0.0001). The shut-off pressure differential between the two centrifugal blood pumps was significant and increased given higher rotation speeds (p < 0.0001). The RotaFlow blood pump has higher maximal flow rate (9.08 +/- 0.01L/min) compared with the CentriMag blood pump (8.37 +/- 0.02L/min) (p < 0.0001). The blood flow rate differential between the two pumps when measured at the same revolutions per minute (RPM) ranged from 1.64L/min to 1.73L/min.
CONCLUSIONS: The results obtained in this experiment demonstrate that the RotaFlow has a higher shut-off pressure (less retrograde flow) and maximal blood flow rate than the CentriMag blood pump. Findings support the conclusion that the RotaFlow disposable pump head has a better mechanical performance than the CentriMag. In addition, the RotaFlow disposable pump is 20-30 times less expensive than the CentriMag.

Entities:  

Mesh:

Year:  2010        PMID: 20212070     DOI: 10.1177/0267659110365366

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  4 in total

1.  Evolution of the extracorporeal life support circuitry.

Authors:  David Palanzo; Feng Qiu; Larry Baer; J Brian Clark; John L Myers; Akif Undar
Journal:  Artif Organs       Date:  2010-11       Impact factor: 3.094

Review 2.  Mechanical circulatory support for end-stage heart failure in repaired and palliated congenital heart disease.

Authors:  Joseph B Clark; Linda B Pauliks; John L Myers; Akif Undar
Journal:  Curr Cardiol Rev       Date:  2011-05

3.  Computational fluid dynamics analysis and experimental hemolytic performance of three clinical centrifugal blood pumps: Revolution, Rotaflow and CentriMag.

Authors:  Dong Han; Joshua L Leibowitz; Lu Han; Shigang Wang; Ge He; Bartley P Griffith; Zhongjun J Wu
Journal:  Med Nov Technol Devices       Date:  2022-06-19

4.  Brain Histopathology of Adult Decedents After Extracorporeal Membrane Oxygenation.

Authors:  Imad R Khan; Yang Gu; Benjamin P George; Laura Malone; Kyle S Conway; Fabienne Francois; Jack Donlon; Nadim Quazi; Ashwin Reddi; Cheng-Ying Ho; Daniel L Herr; Mahlon D Johnson; Gunjan Y Parikh
Journal:  Neurology       Date:  2021-01-20       Impact factor: 9.910

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.