| Literature DB >> 20209661 |
Florian Maire1, Catherine M Lange.
Abstract
We report the formation of unexpected ions during the analyses of a first-generation polyamidoamine dendrimer in negative ion mode using an ion trap equipped with an electrospray ionisation source. These surprising ions corresponded to an increase of 12 m/z units over those expected. The formation of the unexpected ions was dependent on the tuning of the solution flow rate and the capillary high voltage. In addition, measurements of unusual value of the current suggested that a reaction was occurring in the corona plasma. The influence of methanol in this phenomenon was demonstrated by using CD(3)OH in the sample preparation. We propose two structures to explain the observed adduct based on the results of MS(2) experiments and by referring to previous work dealing with 12 m/z units addition. We showed that a corona discharge caused by alterations taking place to the electrospray capillary emitter was the origin of these unexpected ions. Finally, we discuss the mechanism involved in the formation of the ions and we propose means to control such artefacts. 2010 John Wiley & Sons, Ltd.Entities:
Year: 2010 PMID: 20209661 DOI: 10.1002/rcm.4475
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419