Literature DB >> 20208929

Improvement in dynamic range limitation of swept source optical coherence tomography by true logarithmic amplification.

Bin Liu1, Ehsan Azimi, Mark E Brezinski.   

Abstract

We previously demonstrated, with both theoretical and experimental studies, the dynamic range limitation with spectral domain optical coherence tomography (OCT) relative to time domain OCT. A significant portion of this limitation was due to the difference of analog/digital conversion. In this paper, a new method of true logarithmic amplification is discussed theoretically and tested experimentally to increase the dynamic range of a swept source OCT. With the current experimental setup, an increase of the dynamic range by about 6 dB was obtained.

Mesh:

Year:  2010        PMID: 20208929     DOI: 10.1364/JOSAA.27.000404

Source DB:  PubMed          Journal:  J Opt Soc Am A Opt Image Sci Vis        ISSN: 1084-7529            Impact factor:   2.129


  3 in total

1.  Evaluation of optical coherence tomography for the measurement of the effects of activators and anticoagulants on the blood coagulation in vitro.

Authors:  Xiangqun Xu; Jinhai Geng; Gangjun Liu; Zhongping Chen
Journal:  IEEE Trans Biomed Eng       Date:  2013-02-06       Impact factor: 4.538

2.  True logarithmic amplification of frequency clock in SS-OCT for calibration.

Authors:  Bin Liu; Ehsan Azimi; Mark E Brezinski
Journal:  Biomed Opt Express       Date:  2011-05-27       Impact factor: 3.732

3.  Dynamic-range compression and contrast enhancement in swept-source optical coherence tomography systems with a frequency gain compensation amplifier.

Authors:  Shanshan Liang; Xinyu Li; Yao Qin; Jun Zhang
Journal:  J Biomed Opt       Date:  2020-11       Impact factor: 3.170

  3 in total