Non-invasive flow measurement of a rotary pump ventricular assist device using quantitative contrast echocardiography.

Karl Q Schwarz1, Sachin S Parikh, Xucai Chen, David J Farrar, Sherry Steinmetz, Sridevi Ramamurthi, William Hallinan, H Todd Massey, Leway Chen.   

Abstract

BACKGROUND: Many implantable ventricular assist devices (VADs) have no direct measurement of pump output. The aim of this study was to test the hypothesis that quantitative contrast echocardiography can be used to measure VAD output.
METHODS: Contrast-enhanced Doppler velocity-time integral (VTI) was measured in the VAD inlet and outlet cannulae. Doppler flow (Doppler Q=Doppler VTIxcannula area) was compared with measured flow (Q). A total of 130 flow measurements were made (at 6400 and 12,000 rpm).
RESULTS: Doppler Q in the outflow and inflow cannulae showed an excellent correlation with measured Q (outlet Doppler Q=1.0052 xQ+0.048, R2=0.9865; inlet Doppler Q=1.5043 xQ+0.003, R2=0.9904), but inlet Doppler Q was 50% higher. Correcting for the flow profile of the conical inlet tube yielded excellent correlation (inlet Doppler Q=1.0029 xQ+0.002, R2=0.9904).
CONCLUSION: Noninvasive Doppler flow techniques can be used to accurately measure VAD flow, but the flow profile in the cannula needs to be taken into account. Copyright (c) 2010 American Society of Echocardiography. Published by Mosby, Inc. All rights reserved.

Mesh:

Substances:

Year:  2010        PMID: 20206830     DOI: 10.1016/j.echo.2009.12.005

Source DB:  PubMed          Journal:  J Am Soc Echocardiogr        ISSN: 0894-7317            Impact factor:   5.251