An experimental and theoretical study of the effect of sample thickness on the Shore hardness of elastomers.

Adel Siddiqui1, Michael Braden, Mangala P Patel, Sandra Parker.   

Abstract

OBJECTIVE: The Shore hardness test is a quick and convenient way to measure hardness of elastomers. The test specifies that the specimen should be at least 6mm thick; however, published work in the literature indicates that workers often use much thinner samples. The aim of this study was to investigate the effect of thickness on the measured hardness of a range of dental elastomers and develop a theoretical relationship to predict the effect.
METHODS: 7 dental silicone elastomers were selected to give a range of hardness values. Shore A scale hardness measurements were made on standard thickness samples, and those of lower thicknesses. A theoretical analysis was derived, whereby the effect of thickness of the sample on its measured hardness can be predicted and was tested using the results obtained.
RESULTS: The results show reasonable agreement between theoretical and experimental values. The measured Shore hardness increased inversely with thickness, the effect being most pronounced with thin samples. Results ranged from 66.3+/-1.6-80.9+/-0.4 for 1mm thick to 35.0+/-1.7-69.1+/-0.6 for the 6mm thickness. Ratio of the measured hardness of 1mm/6mm ranged from 1.97 for the softest material to 1.17 for the hardest. SIGNIFICANCE: Measurements obtained with a Shore A hardness tester on samples of less than the specified 6mm thickness will give erroneously high values, the error being greater as thickness decreases. The theoretical treatment proposed gives a basis on which data obtained from studies that used thinner specimens can be better analysed. Copyright (c) 2010 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20206378     DOI: 10.1016/j.dental.2010.02.004

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  4 in total

1.  A Novel Silicone Simulation Model for Microvascular Anastomosis.

Authors:  Jessica M Winter; Christian Petropolis
Journal:  Plast Surg (Oakv)       Date:  2021-05-06       Impact factor: 0.558

2.  Optically tracked and 3D printed haptic phantom hand for surgical training system.

Authors:  Johannes Maier; Maximilian Weiherer; Michaela Huber; Christoph Palm
Journal:  Quant Imaging Med Surg       Date:  2020-02

Review 3.  Long-Term Soft Denture Lining Materials.

Authors:  Grzegorz Chladek; Jarosław Żmudzki; Jacek Kasperski
Journal:  Materials (Basel)       Date:  2014-08-12       Impact factor: 3.623

4.  Comparison of the Hardness of Novel Experimental Vinyl Poly Siloxane (VPS) Impression Materials with Commercially Available Ones.

Authors:  Shahab Ud Din; Farooq Ahmad Chaudhary; Bilal Ahmed; Mohammad Khursheed Alam; Sandra Parker; Mangala Patel; Muhammad Qasim Javed
Journal:  Biomed Res Int       Date:  2022-02-09       Impact factor: 3.411

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.