Literature DB >> 20199266

The StarClose vascular closure device in antegrade and retrograde punctures: a single-center experience.

Albert Ho Yuen Chiu1, Simon Richard Coles, Jonathan Tibballs, Sanjay Nadkarni.   

Abstract

PURPOSE: To evaluate the StarClose device and compare its success rates in antegrade and retrograde puncture closures.
METHODS: A retrospective review of all StarClose deployments from April 2005 to July 2007 was performed in a single tertiary referral institution radiology department. In this time period, 143 StarClose devices were deployed in 132 patients (102 men; mean age 68+/-14 years). Of these, 40 (28%) were deployed after antegrade and 103 (72%) after retrograde common femoral arterial punctures. Hospital notes were reviewed to evaluate minor (managed conservatively with compression) and major (requiring surgical intervention) complication rates in the immediate postprocedure period and the following 24 hours. Late complications were also assessed.
RESULTS: There were 11 (7.7%) immediate failures of device deployment: 4/40 (10.0%) antegrade and 7/103 (6.8%) retrograde. Within these 11 punctures, 1 major complication occurred that required surgical retrieval of the device following a retrograde puncture. No other major and 12 (9.1%) minor complications occurred following the 132 successful StarClose deployments. No late complications were seen on clinical or radiological follow-up. The total major complication rate was 0.7% (1/143). The total minor complication rate was 15.4% (22/143): 9/40 (22.5%) following antegrade punctures and 13/103 (12.6%) following retrograde punctures.
CONCLUSION: The StarClose device is associated with a low major complication rate. A higher rate of minor complications was observed following antegrade punctures but all were managed with simple compression. Prospective randomized trials comparing closure devices are needed to evaluate their relative efficacy and safety in antegrade and retrograde punctures.

Entities:  

Mesh:

Year:  2010        PMID: 20199266     DOI: 10.1583/09-2873.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  5 in total

1.  Haemostasis with the FISH Vascular Closure Device after 6 French Transfemoral Accesses in Interventional Radiology: Clinical Results.

Authors:  Peter C Kamusella; Christopher W Lüdtke; Fabian Scheer; Reimer Andresen; Christian Wissgott
Journal:  J Clin Diagn Res       Date:  2017-02-01

2.  Use of ultrasound in the insertion of a vascular closure device: a comparative retrospective study with the standard blind technique.

Authors:  Pierleone Lucatelli; Alessandro Cannavale; Carlo Cirelli; Alessandro d'Adamo; Filippo Maria Salvatori; Fabrizio Fanelli
Journal:  Radiol Med       Date:  2014-08-15       Impact factor: 3.469

3.  Percutaneous Vascular Closure System Based on an Extravascular, Bioabsorbable Polyglycolic Plug (ExoSeal): Results from 1000 Patients.

Authors:  P Kamusella; C Wissgott; T Jahnke; J Brossmann; F Scheer; C W Lüdtke; R Andresen
Journal:  Clin Med Insights Cardiol       Date:  2015-06-10

4.  Access site pseudoaneurysms after endovascular intervention for peripheral arterial diseases.

Authors:  Ahmed Eleshra; Daehwan Kim; Hyung Sub Park; Taeseung Lee
Journal:  Ann Surg Treat Res       Date:  2019-05-29       Impact factor: 1.859

5.  Safety and efficacy of the StarClose vascular closure system following 8-Fr sheath placement for intra-aortic balloon pump: a single-center analysis of 42 consecutive patients.

Authors:  Guang Chu; Wenyi Yang; Guobing Zhang; Zhi Zhang; Shaowen Liu; Baogui Sun; Fang Wang
Journal:  Med Princ Pract       Date:  2014-04-23       Impact factor: 1.927

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.