Perturbations produced by electrospray ionization mass spectrometry in the speciation of aluminium(III)/1,6-dimethyl-4-hydroxy-3-pyridinecarboxylate aqueous solutions.

Valerio B Di Marco1, Luca Raveane, Annalisa Dean, Pietro Traldi.   

Abstract

Electrospray ionization mass spectrometry (ESI-MS) is very often employed to study metal/ligand equilibria in aqueous solution. However, the ionization process can introduce perturbations which affect the speciation results in an unpredictable way. It is necessary to identify these perturbations in order to correctly interpret the ESI-MS speciation results. Aluminium(III)/1,6-dimethyl-4-hydroxy-3-pyridinecarboxylate (DQ716) aqueous solutions at various pH were analysed by ESI-MS, and speciation results were compared with those obtained by equilibrium techniques. Differences observed were both qualitative and quantitative. The ESI-MS spectral changes due to different settings of the following instrumental parameters were analyzed: the solution flow rate (F(S)), the nebulizer gas flow rate (F(G)), the potential applied at the entrance capillary (E(C)), and the temperature of the drying gas (T(G)). The effects produced by F(S) and E(C) on the spectra strongly suggest the key role of surface activity in determining the relative fraction of the ions reaching the detector. The experimental effects of F(S) and T(G) were interpreted considering the presence of at least two reactions in the gas phase and a dimerization occurring in the droplets. These perturbations cannot be generalized because they appear to be chemical system-related and instrument-dependent. Therefore, the identification of perturbations is a required task for any metal-ligand equilibrium study performed by ESI-MS. Our results indicate that perturbations can be identified by evaluating the effects produced in the spectra by a change of instrumental parameters. 2010 John Wiley & Sons, Ltd.

Entities:  

Year:  2010        PMID: 20196194     DOI: 10.1002/rcm.4457

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  1 in total

1.  Improved DFT-based interpretation of ESI-MS of aqueous metal cations.

Authors:  Stuart Bogatko; Emilie Cauët; Paul Geerlings
Journal:  J Am Soc Mass Spectrom       Date:  2013-04-18       Impact factor: 3.109

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.