The effect of vacuum mixing and pre-heating the femoral component on the mechanical properties of the cement mantle.

M Baleani1, E Bialoblocka-Juszczyk, G E Engels, M Viceconti.   

Abstract

We investigated the effect of pre-heating a femoral component on the porosity and strength of bone cement, with or without vacuum mixing used for total hip replacement. Cement mantles were moulded in a manner simulating clinical practice for cemented hip replacement. During polymerisation, the temperature was monitored. Specimens of cement extracted from the mantles underwent bending or fatigue tests, and were examined for porosity. Pre-heating the stem alone significantly increased the mean temperature values measured within the mantle (+14.2 degrees C) (p < 0.001) and reduced the mean curing time (-1.5 min) (p < 0.001). The addition of vacuum mixing modulated the mean rise in the temperature of polymerisation to 11 degrees C and reduced the mean duration of the process by one minute and 50 seconds (p = 0.01 and p < 0.001, respectively). In all cases, the maximum temperature values measured in the mould simulating the femur were < 50 degrees C. The mixing technique and pre-heating the stem slightly increased the static mechanical strength of bone cement. However, the fatigue life of the cement was improved by both vacuum mixing and pre-heating the stem, but was most marked (+ 280 degrees C) when these methods were combined. Pre-heating the stem appears to be an effective way of improving the quality of the cement mantle, which might enhance the long-term performance of bone cement, especially when combined with vacuum mixing.

Mesh:

Substances:

Year:  2010        PMID: 20190321     DOI: 10.1302/0301-620X.92B3.22579

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  1 in total

1.  Biomechanical effects of morphological variations of the cortical wall at the bone-cement interface.

Authors:  Chun-Lin Zhang; Guo-Qi Shen; Kun-Peng Zhu; Dong-Xu Liu
Journal:  J Orthop Surg Res       Date:  2016-07-01       Impact factor: 2.359

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.