Literature DB >> 20173707

The accuracy of the Raypex5 electronic apex locator using stainless-steel hand K-file versus nickel-titanium rotary Mtwo file.

Shiva Sadeghi1, Masoomeh Abolghasemi.   

Abstract

OBJECTIVES: Today many clinicians use both stainless-steel hand K-files and nickel- titanium rotary files during endodontic treatment. It is of great importance for the clinician to have confidence in the accuracy of an apex locator even if these files were used interchangeably. The purpose of this in vitro study was to evaluate the accuracy of the Raypex5 electronic apex locator using stainless-steel hand K-file versus nickel-titanium rotary Mtwo file. STUDY
DESIGN: Twenty straight and single canals of maxillary central teeth were used. Access cavities were prepared; actual working length were determined and compared with electronic working length obtained by means of apex locator Raypex5 using stainless-steel hand K-file and nickel-titanium rotary Mtwo file. Data was analyzed by paired T-test.
RESULTS: There was no significant difference between 15/0.02 stainless-steel hand K- file and 10/0.04 NiTi rotary Mtwo file for the mean differences between actual and electronic working length (p=0.126).
CONCLUSIONS: Under the conditions of this in-vitro study, Raypex5 registered more measurements in acceptable range using 15/0.02 stainless-steel hand K- file and 10/0.04 NiTi rotary Mtwo file. It is possible to use them interchangeably without compromising the working length.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20173707     DOI: 10.4317/medoral.15.e788

Source DB:  PubMed          Journal:  Med Oral Patol Oral Cir Bucal        ISSN: 1698-4447


  4 in total

1.  An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files.

Authors:  Paras Mull Gehlot; Vinutha Manjunath; Mysore Krishnaswamy Manjunath
Journal:  Restor Dent Endod       Date:  2016-01-04

2.  The Precision of Propex Pixi with Different Instruments and Coronal Preflaring Procedures.

Authors:  Inês Ferreira; Ana Cristina Braga; Irene Pina-Vaz
Journal:  Eur Endod J       Date:  2019-07-10

3.  Influence of Root Canal Curvature on the Accuracy of Root ZX Electronic Foramen Locator: An In Vitro Study.

Authors:  Masoud Saatchi; Shiva Iravani; Mehrdad Akhavan Khaleghi; Amin Mortaheb
Journal:  Iran Endod J       Date:  2017

4.  An in vivo comparison of accuracy of two electronic apex locators in determining working length using stainless steel and nickel titanium files.

Authors:  Sanjeeb Chaudhary; Archana Gharti; Bhawana Adhikari
Journal:  Clin Cosmet Investig Dent       Date:  2018-05-22
  4 in total