Evaluation of five routine glucose methods on an Olympus AU5400 analyzer using the CDC hexokinase reference method.

Ke-ke Jia1, Jie Zhang.   

Abstract

BACKGROUND: In China, two common methods are used to detect serum glucose, glucose oxidase (GOD) and hexokinase (HK) methods. We evaluated five commercially available glucose reagent kits with the Center of Disease Control (CDC) HK reference method.
METHODS: Five routine methods were compared with the HK reference method according to Clinical and Laboratory Standards Institute (CLSI) guideline (EP9-A2). These five routine methods were performed on an "open channel" Olympus 5400 analyzer.
RESULTS: All five evaluated methods were closely correlated to the reference method (r > 0.9998). Compared to the HK reference method, the Olympus-HK method and Roche-GOD method showed a very small positive bias (mean +/- SD: 0.08 +/- 0.09 mmol/L and 0.03 +/- 0.08 mmol/L); while the Biosino-HK method, Osborne-GOD method and Baiding-GOD method showed a very small negative bias (mean +/- SD: -0.07 +/- 0.06 mmol/L, -0.09 +/- 0.07 mmol/L and -0.01 +/- 0.04 mmol/L).
CONCLUSIONS: All five reagents we evaluated using matching calibrators showed good correlation to the reference method and small bias. These five glucose reagents are acceptable for use in clinical laboratories.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20170396     DOI: 10.1515/CCLM.2010.018

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  The Reproducibility and Usefulness of Estimated Average Glucose for Hyperglycemia Management during Health Checkups: A Retrospective Cross-Sectional Study.

Authors:  Eun-Hee Nah; Seon Cho; Hyeran Park; Dongwon Noh; Eunjoo Kwon; Han-Ik Cho
Journal:  Healthcare (Basel)       Date:  2022-04-29

2.  Evaluation of the Bioequivalence of Acarbose in Healthy Chinese People.

Authors:  Yan Chen; Fahao Guo; Xin Wang; LuYao Liu; Can Yang; YuQing Xiong; Hong Zhang
Journal:  Clin Pharmacol Drug Dev       Date:  2021-02-19

3.  Analytical performance evaluation of different test systems on serum creatinine assay.

Authors:  Lina He; Jinqi Yu; Guang Han; Di Huang; Liqiao Han; Qiaoxuan Zhang; Yunxiu Wang; Zemin Wan; Xian-Zhang Huang; Yujuan Xiong; Xiaobin Wu
Journal:  J Clin Lab Anal       Date:  2021-12-26       Impact factor: 2.352

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.