| Literature DB >> 20168821 |
C H Palmer, R O Claus, S E Fick.
Abstract
The theory and performance of a two beam differential interferometer for measurement of both surface and bulk waves are described. The system is insensitive to small errors (<1 mm) in focus or in specimen flatness. Both amplitude and phase measurements are demonstrated. The system has been absolutely calibrated and can detect 6 x 10(-4)-A surface wave displacements on glass.Entities:
Year: 1977 PMID: 20168821 DOI: 10.1364/AO.16.001849
Source DB: PubMed Journal: Appl Opt ISSN: 1559-128X Impact factor: 1.980