Accelerated aging, natural aging, and small punch testing of gamma-air sterilized polycarbonate urethane acetabular components.

S M Kurtz1, R Siskey, M Reitman.   

Abstract

The objectives of this study were three-fold: (1) to determine the applicability of the small punch test to characterize Bionate 80A polycarbonate urethane (PCU) acetabular implants; (2) to evaluate the susceptibility of PCU acetabular implants to exhibit degradation of mechanical behavior following gamma irradiation in air and accelerated aging; and (3) to compare the oxidation of gamma-air sterilized PCU following accelerated aging and 5 years of natural shelf aging. In addition to attenuated total reflectance-Fourier transform infrared spectroscopy, we also adapted a miniature specimen mechanical test, the small punch test, for the deformable PCU cups. Accelerated aging was performed using ASTM F2003, a standard test that represents a severe oxidative challenge. The results of this study suggest that the small punch test is sufficiently sensitive and reproducible to discriminate slight differences in the large-deformation mechanical behavior of Bionate 80A following accelerated aging. The gamma-air sterilized PCU had a reduction of 9% in ultimate load after aging. Five years of shelf aging had little effect on the mechanical properties of the PCU. Overall, our findings suggest that the Bionate 80A material has greater oxidative stability than ultra-high molecular weight polyethylene following gamma irradiation in air and exposure to a severe oxidative challenge. (c) 2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20166119     DOI: 10.1002/jbm.b.31601

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  6 in total

1.  The MOVE-C Cervical Artificial Disc - Design, Materials, Mechanical Safety.

Authors:  Annette Kienle; Nicolas Graf; Carina Krais; Hans-Joachim Wilke
Journal:  Med Devices (Auckl)       Date:  2020-09-25

2.  The LP-ESP(®) lumbar disc prosthesis with 6 degrees of freedom: development and 7 years of clinical experience.

Authors:  Jean-Yves Lazennec; Alain Aaron; Adrien Brusson; Jean-Patrick Rakover; Marc-Antoine Rousseau
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-01-11

3.  Bionate Biocompatibility: In Vivo Study in Rabbits.

Authors:  Amparo Vanaclocha-Saiz; Vicente Vanaclocha; Carlos Atienza; Pablo Jorda-Gomez; Víctor Primo-Capella; Carlos Barrios; Leyre Vanaclocha
Journal:  ACS Omega       Date:  2022-08-19

4.  Study of the polycarbonate-urethane/metal contact in different positions during gait cycle.

Authors:  Sergio Gabarre; Antonio Herrera; Jesús Mateo; Elena Ibarz; Antonio Lobo-Escolar; Luis Gracia
Journal:  Biomed Res Int       Date:  2014-08-27       Impact factor: 3.411

5.  The innovative viscoelastic CP ESP cervical disk prosthesis with six degrees of freedom: biomechanical concepts, development program and preliminary clinical experience.

Authors:  Jean-Yves Lazennec; Alain Aaron; Olivier Ricart; Jean Patrick Rakover
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-09-04

Review 6.  A critical review of fibrous polyurethane-based vascular tissue engineering scaffolds.

Authors:  Reza Rahbarghazi; Soodabeh Davaran; Sonia Fathi-Karkan; Behnaz Banimohamad-Shotorbani; Sepideh Saghati
Journal:  J Biol Eng       Date:  2022-03-24       Impact factor: 4.355

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.