Qiong Zhang1, Sathyanarayanan Srinivasan, Ying Wu, Siraj Natah, Jeff F Dunn.
Abstract
Broadband near-infrared spectroscopy (bNIRS) is a powerful non-invasive technique for the measurement of hemoglobin. bNIRS systems are relatively simple to construct compared with many near-infrared instruments since they operate on the principle of continuous wave. The advantage of the broadband method is the capacity to model the spectra and to use "the second differential method" to quantify deoxyhemoglobin (HHb). An "anoxia pulse" method can be applied to quantify total haemoglobin (tHb) and tissue oxygen saturation (S(t)O(2)). A disadvantage is that this calibration method is not suitable for application in humans. In this study, we compared the "anoxia pulse" method with "graded hypoxia" method, which can be applied for human studies, to quantify tHb and S(t)O(2). The values obtained with the two methods were respectively (tHb=47.8+/-2.8 and 49.4+/-7.7 microM, mean+/-S.D., n=8) and (S(t)O(2)=72.8+/-3.7% and 73.2+/-5.7%, mean+/-S.D., n=8). There was no significant difference (p<0.05) between the two methods, indicating that the graded hypoxia method could be used for quantification of tHb and S(t)O(2) in human subjects. Copyright (c) 2010 Elsevier B.V. All rights reserved.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20156483 DOI: 10.1016/j.jneumeth.2010.01.037
Source DB: PubMed Journal: J Neurosci Methods ISSN: 0165-0270 Impact factor: 2.390