A Chen1, C Willis-Owen, K Akhtar, S Kamineni.
Abstract
We describe a case of an ASNIS III 5.0 mm partially threaded cannulated screw unthreading itself as it was being inserted during fracture fixation of a humerus. The majority of complications associated with cannulated screws involve breaking of the screw, as opposed to unthreading. We believe that the self tapping design of the screw, in combination with the cannulated design, creates a potential area of weakness when used on hard bone.Entities:
Year: 2010 PMID: 20150983 PMCID: PMC2820442 DOI: 10.1186/1757-1626-3-9
Source DB: PubMed Journal: Cases J ISSN: 1757-1626
Figure 1Picture of Asnis III 5.0 mm Cannulated Screw.
Figure 2Image of unthreaded screw at time of surgery.
Figure 3Image of unthreaded screw 3 weeks following surgery.