Retrieval analysis of femoral zirconium components in total knee arthroplasty: preliminary results.

Thomas J Heyse1, Jack Davis, Steven B Haas, Dan X Chen, Timothy M Wright, Richard S Laskin.   

Abstract

Safety of oxidized zirconium (OxZr) in total knee arthroplasty (TKA) has been supported by biomechanical, clinical, and radiologic data. Retrieved OxZr femoral components and corresponding polyethylene (PE) inserts were examined to rule out patterns leading to early failure. Sixteen retrieved TKA with an OxZr femoral component were included. The PE inlays were analyzed applying an established scoring system for wear and surface damage. Femoral components were screened for scratching, pitting, delamination, and striations. The prostheses were in situ for 16.4 ± 11.9 months. The average wear and damage score for the tibial PE inserts was 36.0 ± 12.7. The average score in the visual analysis of the OxZr femoral components was 1.3 ± 1.3. The data presented in this study did not show major wear of the PE in TKA using OxZr components. In this cohort, there were no failures directly related to this implant.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20149583     DOI: 10.1016/j.arth.2009.11.024

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  5 in total

1.  The 2012 Mark Coventry award: a retrieval analysis of high flexion versus posterior-stabilized tibial inserts.

Authors:  Nicholas R Paterson; Matthew G Teeter; Steven J MacDonald; Richard W McCalden; Douglas D R Naudie
Journal:  Clin Orthop Relat Res       Date:  2013-01       Impact factor: 4.176

Review 2.  Current status and future potential of wear-resistant coatings and articulating surfaces for hip and knee implants.

Authors:  Charlotte Skjöldebrand; Joanne L Tipper; Peter Hatto; Michael Bryant; Richard M Hall; Cecilia Persson
Journal:  Mater Today Bio       Date:  2022-04-30

3.  Oxidized zirconium versus cobalt-chromium in TKA: profilometry of retrieved femoral components.

Authors:  Thomas J Heyse; Marcella E Elpers; Danyal H Nawabi; Timothy M Wright; Steven B Haas
Journal:  Clin Orthop Relat Res       Date:  2014-01       Impact factor: 4.176

4.  Analysis of Carbon Fiber Reinforced PEEK Hinge Mechanism Articulation Components in a Rotating Hinge Knee Design: A Comparison of In Vitro and Retrieval Findings.

Authors:  Ronja A Schierjott; Alexander Giurea; Hans-Joachim Neuhaus; Jens Schwiesau; Andreas M Pfaff; Sandra Utzschneider; Gianluca Tozzi; Thomas M Grupp
Journal:  Biomed Res Int       Date:  2016-12-22       Impact factor: 3.411

5.  High wear resistance of femoral components coated with titanium nitride: a retrieval analysis.

Authors:  Christian Fabry; Carmen Zietz; Axel Baumann; Reinhard Ehall; Rainer Bader
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-05-20       Impact factor: 4.342

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.