Literature DB >> 20146337

Complete stent dislodgement after successful implantation--a rare case.

Hwa Wooi Gan1, Anuj Bhasin, Chiung-Jen Wu.   

Abstract

A successfully deployed stent is normally stable in position and unlikely to be dislodged from the coronary artery. We report a case of complete stent dislodgement after successful implantation in a patient with recent anterior myocardial infarction (AMI). Coronary angulation, a highly calcified coronary artery, an under-deployed stent, and the process of rewiring contributed to this unexpected event. Fortunately, no unwanted complication occurred as a consequence of this event. Copyright 2010 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20146337     DOI: 10.1002/ccd.22368

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692