| Literature DB >> 20141152 |
M Bercovici1, G V Kaigala, C J Backhouse, J G Santiago.
Abstract
We present a novel method for fluorescence-based indirect detection of analytes and demonstrate its use for label-free detection of chemical toxins in a hand-held device. We fluorescently label a mixture of low-concentration carrier ampholytes and introduce it into an isotachophoresis (ITP) separation. The carrier ampholytes provide a large number of fluorescent species with a wide range of closely spaced effective electrophoretic mobilities. Analytes focus under ITP and displace subsets of these carrier ampholytes. The analytes are detected indirectly and quantified by analyzing the gaps in the fluorescent ampholyte signal. The large number (on the order of 1000) of carrier ampholytes enables detection of a wide range of analytes, requiring little a priori knowledge of their electrophoretic properties. We discuss the principles of the technique and demonstrate its use in the detection of various analytes using a standard microscope system. We then present the integration of the technique into a self-contained hand-held device and demonstrate detection of chemical toxins (2-nitrophenol and 2,4,6-trichlorophenol) in tap water, with no sample preparation steps.Entities:
Year: 2010 PMID: 20141152 DOI: 10.1021/ac902526g
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986