Measurement of anterior chamber volume with rotating scheimpflug camera and anterior segment optical coherence tomography.

Jing Fu1, Shu-ning Li, Xiao-zhen Wang, Ge-wei Wu, Da-peng Mu, Jian Wang, Ning-li Wang.   

Abstract

BACKGROUND: Measurement of anterior segment parameters plays an important role in diagnosis and treatment of glaucoma. The objective of this study was to evaluate the repeatability and reproducibility of anterior chamber volume (ACV) measurements with rotating scheimpflug camera (RSC) and to examine agreement with anterior segment optical coherence tomography (AS-OCT).
METHODS: Thirty nine healthy normal subjects were recruited from the Eye Center of Tongren Hospital. ACV was measured using RSC and AS-OCT in a randomly selected eye for each subject. For RSC measurements, both automatic and manual ACV measurements and 2 independent operators' ACV measurements were obtained. All subjects were invited for 3 visits within a week to evaluate repeatability and reproducibility of ACV measurement by RSC. Agreement was evaluated between RSC and AS-OCT.
RESULTS: Good repeatability and reproducibility were found for both automatic and manual ACV measurements obtained by RSC. For intrasession repeatability, coefficient of variation (CVw) and intraclass correlation coefficient (ICC) values for automatic were 3.52% and 0.98; the values for manual were 3.44% and 0.97, respectively. For intersession reproducibility, the respective CVw and ICC values were 3.96% and 0.96. Good agreement was also found in 2 operators for both automatic and manual ACV measurements; nevertheless, poor agreement was found between RSC and AS-OCT (95% confidence interval (CI) for agreement of automatic RSC measurement versus AS-OCT were -96.3 to 72.8 microl and 95% CI for agreement of manual RSC measurement versus AS-OCT were between -41.7 to 10.1 microl).
CONCLUSIONS: Both RSC automatic and manual ACV measurements showed good repeatability and reproducibility, and showed comparable agreement between 2 independent operators, but poor agreement was found between RSC and AS-OCT.

Entities:  

Mesh:

Year:  2010        PMID: 20137371

Source DB:  PubMed          Journal:  Chin Med J (Engl)        ISSN: 0366-6999            Impact factor:   2.628


  3 in total

1.  Comparison of the anterior ocular segment measurements using swept-source optical coherent tomography and a scanning peripheral anterior chamber depth analyzer.

Authors:  Toshie Furuya; Fumihiko Mabuchi; Tatsuya Chiba; Satoshi Kogure; Shigeo Tsukahara; Kenji Kashiwagi
Journal:  Jpn J Ophthalmol       Date:  2011-08-10       Impact factor: 2.447

2.  Objective evaluation of the changes in the crystalline lens during accommodation in young and presbyopic populations using Pentacam HR system.

Authors:  Yao Ni; Xia-Lin Liu; Ming-Xing Wu; Ying Lin; Yu-Ying Sun; Chang He; Yi-Zhi Liu
Journal:  Int J Ophthalmol       Date:  2011-12-18       Impact factor: 1.779

3.  Comparison of Anterior Segment Measurements Obtained by Aladdin Optical Biometer and Sirius Corneal Topography.

Authors:  Onur Polat; Zeki Baysal; Serkan Özcan; Sibel İnan; Ümit Übeyt İnan
Journal:  Turk J Ophthalmol       Date:  2016-12-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.