| Literature DB >> 20133064 |
Esben Thormann1, Torbjörn Pettersson, John Kettle, Per M Claesson.
Abstract
A phase shift between the oscillatory motion and drive motion of an AFM-cantilever used for tapping mode AFM imaging can be related to adhesive and elastic properties of surface layers. In this study it was studied how optimal contrast between hard and soft surface layers can be achieved while minimizing the surface damage. This was investigated by performing classical force-distance measurements while driving the cantilever as in tapping mode imaging. The amplitude and phase response as a function of the average tip-surface separation was recorded. Five different cantilevers with a wide range of spring constants and four different tapping amplitudes were investigated and compared. Based on these experiments it is concluded that too stiff cantilever, high free tapping amplitude and low amplitude set point value often lead to surface damage, while too low spring constant and low free tapping amplitude result in poor phase image contrast. Intermediate values where little surface damage and significant image contrast are obtained were identified. In all cases it was observed that the best image contrast was obtained when the amplitude set point was chosen such that the amplitude during imaging was reduced to approximately 50% of the free amplitude. (c) 2010 Elsevier B.V. All rights reserved.Year: 2010 PMID: 20133064 DOI: 10.1016/j.ultramic.2010.01.002
Source DB: PubMed Journal: Ultramicroscopy ISSN: 0304-3991 Impact factor: 2.689