Model-based ultrasound temperature visualization during and following HIFU exposure.

Guoliang Ye1, Penny Probert Smith, J Alison Noble.   

Abstract

This paper describes the application of signal processing techniques to improve the robustness of ultrasound feedback for displaying changes in temperature distribution in treatment using high-intensity focused ultrasound (HIFU), especially at the low signal-to-noise ratios that might be expected in in vivo abdominal treatment. Temperature estimation is based on the local displacements in ultrasound images taken during HIFU treatment, and a method to improve robustness to outliers is introduced. The main contribution of the paper is in the application of a Kalman filter, a statistical signal processing technique, which uses a simple analytical temperature model of heat dispersion to improve the temperature estimation from the ultrasound measurements during and after HIFU exposure. To reduce the sensitivity of the method to previous assumptions on the material homogeneity and signal-to-noise ratio, an adaptive form is introduced. The method is illustrated using data from HIFU exposure of ex vivo bovine liver. A particular advantage of the stability it introduces is that the temperature can be visualized not only in the intervals between HIFU exposure but also, for some configurations, during the exposure itself. 2010 World Federation for Ultrasound in Medicine & Biology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20113861     DOI: 10.1016/j.ultrasmedbio.2009.10.001

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  4 in total

Review 1.  Focused Ultrasound for Neuromodulation.

Authors:  David P Darrow
Journal:  Neurotherapeutics       Date:  2019-01       Impact factor: 7.620

Review 2.  Thermometry and ablation monitoring with ultrasound.

Authors:  Matthew A Lewis; Robert M Staruch; Rajiv Chopra
Journal:  Int J Hyperthermia       Date:  2015-03-10       Impact factor: 3.914

3.  Calibration and Evaluation of Ultrasound Thermography Using Infrared Imaging.

Authors:  Yi-Sing Hsiao; Cheri X Deng
Journal:  Ultrasound Med Biol       Date:  2015-11-05       Impact factor: 2.998

4.  An approach for the visualization of temperature distribution in tissues according to changes in ultrasonic backscattered energy.

Authors:  Jingjing Xia; Qiang Li; Hao-Li Liu; Wen-Shiang Chen; Po-Hsiang Tsui
Journal:  Comput Math Methods Med       Date:  2013-10-24       Impact factor: 2.238

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.