Literature DB >> 20113264

Online conductivity monitoring of dialysis adequacy versus Kt/V derived from urea reduction ratio: a prospective study from a Saudi Center.

Khalid Al Saran1, Alaa Sabry, Mamdouh Abdulghafour, Ahmed Yehia.   

Abstract

INTRODUCTION: Adequate delivered dose of solute removal (as assessed by urea reduction and calculation of Kt/V) is an important determinant of clinical outcome in chronic hemodialysis (HD) patients. This requires both prescription of an adequate dose of HD and regular assessment that the delivered treatments are also adequate. Online conductivity monitoring (OCM) using sodium flux as a surrogate for urea allows for the repeated non-invasive measurement of Kt/V on each HD treatment.
METHODS: We prospectively studied 17 (9 males, 8 females) established chronic HD patients over an eight-week period (408 treatments). A pre- and post-dialyzer measurement of the conductivity is performed by two mutually independent temperature-compensated conductivity cells equipped with Fresenius 4008 S(R) dialysis machines. Urea reduction was measured once a week by a single-pool calculation using immediate post-treatment sampling. No changes were made to any of the dialysis prescriptions over the study period. Values of calculated Kt/V and simultaneously obtained online Kt/V were compared.
RESULTS: There was a statistically significant difference between calculated Kt/V and online Kt/V over the study period. The mean calculated Kt/V was 1.37 +/- 0.09, and mean online Kt/V 1.02 +/- 0.15 (p = 0.000). Calculated Kt/V >or= 1.2 was achieved in all our patients, while online Kt/V >or= 1.2 was achieved in only 17.64%. Yet there was moderate correlation between calculated Kt/V and online Kt/V (r(2) = 0.48).
CONCLUSIONS: Online conductivity monitoring (OCM) results underestimates dialysis efficiency compared to calculated Kt/V readings. This difference has to be considered when applying Kt/V to clinical practice.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20113264     DOI: 10.3109/08860220903367486

Source DB:  PubMed          Journal:  Ren Fail        ISSN: 0886-022X            Impact factor:   2.606


  2 in total

1.  Comparison between different methods of calculating Kt/V as the marker of adequacy of dialysis.

Authors:  Sajad Ahmad; Irfan Elahi; Muhammad Anees
Journal:  Pak J Med Sci       Date:  2022 Jan-Feb       Impact factor: 1.088

2.  Monitoring dialysis adequacy: history and current practice.

Authors:  Linda Ding; James Johnston; Maury N Pinsk
Journal:  Pediatr Nephrol       Date:  2021-01-05       Impact factor: 3.714

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.