Literature DB >> 20113257

Is heparin plasma suitable for the determination of B-type natriuretic peptide on the Beckman-Coulter Access 2?

Humberto Santos1, Bruno Cauliez, Christophe Tron, Valery Brunel, Alain Lavoinne.   

Abstract

BACKGROUND: The use of heparin as an alternative to EDTA in the production of plasma samples is of particular interest for B-type natriuretic peptide (BNP) measurements. Lithium heparin is now widely used for the determination of biochemical parameters, including cardiac markers. The goal of this study was to determine the feasibility of measuring BNP using heparin plasma instead of EDTA plasma with the Access 2 system (Beckman-Coulter).
METHODS: BNP was determined in heparin plasma and EDTA plasma from 24 patients within 1 h of blood collection. Additional measurements were performed with heparin plasma, every hour for the first 4 h, and then 8 h after the collection of blood that was stored at room temperature.
RESULTS: At H(0), the observed BNP concentrations in heparin plasma were much higher (mean values 65% higher) than those in EDTA plasma. Using predetermined thresholds, this difference would lead to 30% discordance between samples in heparin and EDTA. BNP stability decreased over time in heparin plasma: immunoreactivity decreased approximately by 30% during the first 2 h and by 60% after 8 h.
CONCLUSIONS: Heparin plasma does not seem to be a suitable alternative to EDTA plasma for measurement of BNP using the Access 2 system, even if measurements are performed immediately after blood sampling.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20113257     DOI: 10.1515/CCLM.2010.069

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  The standards for reporting diagnostic accuracy studies 2015 update: is there a missing link to the triumvirate?

Authors:  Alan H B Wu; Robert H Christenson
Journal:  Ann Transl Med       Date:  2016-02

2.  Factors associated with serum B-type natriuretic peptide in infants with single ventricles.

Authors:  Ryan J Butts; Victor Zak; Daphne Hsu; James Cnota; Steven D Colan; David Hehir; Paul Kantor; Jami C Levine; Renee Margossian; Marc Richmond; Anita Szwast; Derek Williams; Richard Williams; Andrew M Atz
Journal:  Pediatr Cardiol       Date:  2014-02-13       Impact factor: 1.655

3.  Analytical performance of a single epitope B-type natriuretic peptide sandwich immunoassay on the Minicare platform for point-of-care diagnostics.

Authors:  Alexander van Reenen; Mario Berger; Emmanuel Moreau; Edwin Bekx; Tom Bruinink; Danielle Kemper; Lian van Lippen; Jos Weusten; Anita Mrakovic; Etienne Michielsen; Joost Vissers; Femke de Theije; Jeroen Nieuwenhuis; Veronique Semjonow; Johannes Mair
Journal:  Pract Lab Med       Date:  2019-03-17
  3 in total