Literature DB >> 20109421

Intermittent atrial undersensing in single-lead VDD pacemakers in patients with bradycardia-sensitive repolarization: a possible mechanism for ventricular arrhythmia.

Víctor Palanca1, Alejandro Navarro, Javier Jiménez, Aurelio Quesada, Salvador Morell, José Roda.   

Abstract

Intermittent atrial undersensing in patients with single-lead VDD pacemakers is regarded as being of limited clinical significance. Nevertheless, in patients with bradycardia-mediated repolarization abnormalities, undersensing could result in RR-interval oscillations due to changes in pacemaker mode and this could act as a trigger for the initiation of torsades de pointes. We report our findings in three patients, which demonstrate the functioning of this trigger mechanism for ventricular arrhythmias.

Entities:  

Mesh:

Year:  2010        PMID: 20109421     DOI: 10.1016/s1885-5857(10)70042-6

Source DB:  PubMed          Journal:  Rev Esp Cardiol        ISSN: 0300-8932            Impact factor:   4.753


  1 in total

1.  Polymorphic ventricular tachycardia due to change in pacemaker programming.

Authors:  Ihab Elsokkari; Amir Abdelwahab; Ratika Parkash
Journal:  HeartRhythm Case Rep       Date:  2017-02-20
  1 in total