Development and validation of a blinded hybrid device according to the European Hypertension Society protocol: Nissei DM-3000.

F Tasker1, A De Greeff, A H Shennan.   

Abstract

Both automated and auscultatory blood pressure (BP) devices have their strengths and accuracy limitations. Hybrid devices, such as the Nissei DM-3000, are mercury free and provide both automated and auscultatory measurement modes. The aim of this study was to validate all measurement modes of the Nissei DM-3000 device according to the European Society of Hypertension (ESH) protocol, as well as to develop and validate a 'blinded' auscultatory measurement mode. Different measurement modes were developed and evaluated in separate studies. Nine sequential same-arm BP measurements were taken alternating between simultaneous mercury sphygmomanometer readings and the device. The latter seven measurements were analysed according to the requirements of the ESH protocol. All measurement modes of the device passed the ESH protocol. The blinded mode achieved the best results with a mean difference+/-s.d. of -0.1+/-2.6 and 0.04+/-2.4 mm Hg for systolic BP (SBP) and diastolic BP (DBP), respectively. The most accurate auscultatory measurement results were obtained with a deflation rate of 2.5 mm Hg s(-1) achieving a mean difference+/-s.d. of -0.6+/-4.4 (for SBP) and -1.4+/-2.8 mm Hg (for DBP). The automated mode achieved a mean difference+/-s.d. of -0.8+/-6.0 (SBP) and 0.8+/-4.8 mm Hg (DBP). The Nissei DM-3000 device is a suitable replacement for the mercury sphygmomanometer.

Entities:  

Mesh:

Year:  2010        PMID: 20107491     DOI: 10.1038/jhh.2009.113

Source DB:  PubMed          Journal:  J Hum Hypertens        ISSN: 0950-9240            Impact factor:   3.012


  3 in total

1.  From mercury sphygmomanometer to electric device on blood pressure measurement: correspondence of Minamata Convention on Mercury.

Authors:  Kei Asayama; Takayoshi Ohkubo; Satoshi Hoshide; Kazuomi Kario; Yusuke Ohya; Hiromi Rakugi; Satoshi Umemura
Journal:  Hypertens Res       Date:  2016-01-28       Impact factor: 3.872

Review 2.  Blood pressure measurement: clinic, home, ambulatory, and beyond.

Authors:  Paul E Drawz; Mohamed Abdalla; Mahboob Rahman
Journal:  Am J Kidney Dis       Date:  2012-04-21       Impact factor: 8.860

3.  Factors influencing the accuracy of non-invasive blood pressure measurements in patients admitted for cardiogenic shock.

Authors:  Jitka Seidlerová; Pavlína Tůmová; Richard Rokyta; Milan Hromadka
Journal:  BMC Cardiovasc Disord       Date:  2019-06-18       Impact factor: 2.298

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.